Agilent Technologies 1100 Series Reference Manual page 136

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector
2
If not already in this position, move the holmium oxide
filter down.
4
Carefully slide the assembly to the right into the flow cell
compartment by pushing it from the holmium oxide filter
side.
136
3
Use the 1.5-mm hex key to loosen the setscrew that
secures the cell support assembly.
5
Remove the cell support assembly completely.
1100 Series DAD and MWD Reference Manual