Repairing the Detector

Repairing the High Pressure Flow Cell

When required
Tools required
Parts required
Preparations for this procedure:
Turn off the flow.
Remove the front cover.
Remove the flow cell, see
page 105.
112
If the flow cell needs repair due to leaks or contaminations (reduced light
throughput)
Two 1/4 inch wrenches for capillary connections
hexagonal key 4 mm
Tooth picks
For parts see
"Exchanging a Flow Cell"
"High Pressure Flow Cell"
1 Remove the window assembly (1) and the gasket (2) from
the cell body.
on
1
2
1100 Series DAD and MWD Reference Manual
on page 190
2
1