Replacing Status Light Pipe - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector

Replacing Status Light Pipe

When required
Tools required
Parts required
Preparations for this procedure:
Remove the front cover and top cover, see
the Top Cover and Foam"
Next steps:
Replace the top cover, see
Replace the detector into the stack and reconnect the cables and capillaries.
Turn on the detector.
170
If part was broken or removed
Screwdriver POZI 1 PT3
Status light pipe 5041-8384
"Removing
on page 140.
"Replacing the Foam and the Top Cover"
1 The status light pipe is clipped into the top cover.
on page 175.
1100 Series DAD and MWD Reference Manual