Agilent Technologies 1100 Series Reference Manual page 159

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2 Pull the filter motor off the shaft. Keep the spring and the
lever fixture in a safe place. Remove the filter lever on the
other side.
4 Use the 1.5-mm hexagonal key to loosen the setscrew that
secures the cell support assembly.
1100 Series DAD and MWD Reference Manual
3 Use the 1.5-mm hexagonal key to loosen the setscrew that
secures the source lens assembly.
5 Carefully slide the assembly to the right by pushing it from
the deuterium lamp side with a hexagonal wrench.
Repairing the Detector
4
159