Repairing the Detector
5 Unscrew the fixing screw and unwrap the inlet capillary
from the grove in the flow cell body.
Screw
Inlet capillary
with heat
7 Bend the caillary again by 90° as shown below.
116
90°
6 Take the new inlet capillary and bend it 90° about 35 mm
from its end.
35 mm
90°
8 Insert the capillary into the hole between fixing screw and
the inlet fitting.
1100 Series DAD and MWD Reference Manual