Cleaning or Exchanging the Holmium Oxide Filter

When required
Tools required
Parts required
See also
N O T E
Preparations for this procedure:
Turn off the flow.
Remove the front cover.
Remove the flow cell, see
page 105.
1100 Series DAD and MWD Reference Manual
If holmium oxide filter is contaminated
Screwdriver POZI 1 PT3
Screwdriver flat blade
Two 1/4 inch wrenches for capillary connections
A pair of tweezers
Holmium oxide filter 79880-22711
"Declaration of Conformity for HOX2 Filter"
"Exchanging a Flow Cell"
on page 310.
1
Unscrew the six screws and remove the flow cell cover.
on
Repairing the Detector
4
127