Agilent Technologies 1100 series Reference Manual
Agilent Technologies 1100 series Reference Manual

Agilent Technologies 1100 series Reference Manual

Binary pump
Hide thumbs Also See for 1100 series:
Table of Contents

Advertisement

Agilent G1312A Bin Pump
Asset # 304023
Agilent 1100 Series
Binary Pump
Reference Manual
sa

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 1100 series

  • Page 1 Agilent 1100 Series Binary Pump Agilent G1312A Bin Pump Asset # 304023 Reference Manual...
  • Page 2 270. Reproduction, adaption, change without notice. or translation without Warning Symbols Used prior written permission Agilent Technologies In This Book is prohibited, except as makes no warranty of allowed under the any kind with regard to copyright laws.
  • Page 3 Agilent 1100 Series Binary Pump Reference Manual...
  • Page 4 In This Book This manual contains technical reference information about the Agilent 1100 Series binary pump. The manual describes the following: • installation, • optimizing performance, • troubleshooting, • repairing, • parts and materials, • introduction and theory of operation, •...
  • Page 5: Table Of Contents

    Contents 1 Installing the Pump How to install the binary pump Site Requirements 14 Unpacking the Binary Pump 17 Optimizing the Stack Configuration 19 Installing the Binary Pump 21 Flow Connections of the Binary Pump with Solvent Selection Valve 24 Flow Connections of the Binary Pump without Solvent Selection Valve 27 Priming and Purging the System 30...
  • Page 6 Contents functions Status Indicators Power Supply Indicator 49 Instrument Status Indicator 49 Error Messages Timeout 51 Shutdown 52 Remote Timeout 53 Synchronization Lost 54 Leak 55 Leak Sensor Open 56 Leak Sensor Short 57 Compensation Sensor Open 58 Compensation Sensor Short 59 Fan Failed 60 Open Cover 61 Restart Without Cover 62...
  • Page 7 Contents Encoder Missing 75 Inlet-Valve Missing 76 Servo Restart Failed 77 Pump Head Missing 78 Index Limit 79 Index Adjustment 80 Index Missing 81 Stroke Length 82 Initialization Failed 83 Wait Timeout 84 Pressure Test Running the Pressure Test 87 Evaluating the Results 89 Leak Test Description 92...
  • Page 8 Contents Simple Repair Procedures Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve 109 Exchanging the Outlet Ball Valve Sieve or the Complete Valve 112 Exchanging the Purge Valve Frit or the Purge Valve 114 Exchanging the Solvent Selection Valve 117 Removing and Disassembling the Pump Head Assembly 119 Exchanging the Pump Seals and Seal Wear-in Procedure 121 Exchanging the Plungers 124...
  • Page 9 Contents 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials Control Module(B-Version) 167 Solvent Cabinet 168 Bottle-Head Assembly 169 Hydraulic Path with Solvent Selection Valve 170 Hydraulic Path without Solvent Selection Valve 172 Cover Parts 174 Sheet Metal Kit 175 Foam Parts 176 Power and Status Light Pipes 177...
  • Page 10 Early Maintenance Feedback (EMF) 213 The Electronics 215 Firmware Description 221 Optional Interface Boards 223 Agilent 1100 Series Interfaces 225 Setting the 8-bit Configuration Switch 230 The Main Power Supply Assembly 235 7 Control Module Screens for the Binary Pump...
  • Page 11 Contents Warranty Statement Safety Information Lithium Batteries Information Radio Interference Sound Emission Solvent Information Agilent Technologies on Internet...
  • Page 12 Contents...
  • Page 13: Installing The Pump

    Installing the Pump How to install the binary pump...
  • Page 14: Site Requirements

    Installing the Pump Site Requirements A suitable environment is important to ensure optimum performance of the binary pump. Power Consideration The binary pump power supply has wide ranging capability (see Table 1 on page 16). It accepts any line voltage in the range described in the above mentioned table.
  • Page 15 (1.0 inches) of space on either side and approximately 8 cm (3.1 inches) in the rear for the circulation of air and electric connections. If the bench should carry a complete Agilent 1100 Series system, make sure that the bench is designed to carry the weight of all the modules.
  • Page 16 Installing the Pump Site Requirements Table 1 Physical Specifications Type Specification Comments Weight 15.5 kg (34 lbs) Dimensions 180 × 345 × 435 mm (height × weight × depth) (7 × 13.5 × 17 inches) Line voltage ± Wide-ranging capability 100 –...
  • Page 17: Unpacking The Binary Pump

    Ensure all parts and materials have been delivered with the binary pump. The delivery checklist is shown in Table 2. To aid in parts identification, please see Chapter 5 “Parts and Materials”. Please report missing or damaged parts to your local Agilent Technologies sales and service office. Table 2 Binary Pump Checklist...
  • Page 18 Installing the Pump Unpacking the Binary Pump Table 2 Binary Pump Checklist, continued Description Quantity CAN cable, 1 m Remote cable As ordered Signal cable As ordered Reference Manual Accessory kit (see Table 3) 1 4 bottles when binary pump is equipped with a solvent selection valve 4 bottle-head assemblies when binary pump is equipped with a solvent selection valve Accessory Kit Contents - Binary Pump Table 3...
  • Page 19: Optimizing The Stack Configuration

    Optimizing the Stack Configuration Optimizing the Stack Configuration If your binary pump is part of a complete 1100 series system, you can ensure optimum performance by limiting the configuration of the system stack to the following configuration. This configuration optimizes the system flow path, ensuring minimum delay volume.
  • Page 20 Installing the Pump Optimizing the Stack Configuration Figure 2 Recommended Stack Configuration (Rear View) Pressure output to recorder, for PN see page 188 CAN Bus cable to handheld controller G1323-81600 CAN Bus cable for inter module communication 5181-1516 (0.5m) 5161-1519 (1.0m) Analog signal to AC power recorder,for...
  • Page 21: Installing The Binary Pump

    Installing the Pump Installing the Binary Pump Installing the Binary Pump Preparations Locate bench space Provide power connections Unpack the pump Parts required Pump Power cord, for other cables see text below and “Cable Overview” on page 188 ChemStattion and/or Control module G1323A/B 1 Place the binary pump on the bench in a horizontal position.
  • Page 22 N O T E In an Agilent 1100 Series system, the individual modules are connected through CAN cables. The Agilent 1100 Series vacuum degasser is an exception . A vacuum degasser can be connected via the APG remote connector to the other modules of the stack.
  • Page 23 Installing the Pump Installing the Binary Pump 6 Connect the capillary, solvent tubes and waste tubings (see “Flow Connections of the Binary Pump with Solvent Selection Valve” on page 24 or “Flow Connections of the Binary Pump without Solvent Selection Valve” on page 27).
  • Page 24: Flow Connections Of The Binary Pump With Solvent Selection Valve

    Installing the Pump Flow Connections of the Binary Pump with Solvent Selection Valve Flow Connections of the Binary Pump with Solvent Selection Valve WAR NI N G When opening capillary or tube fittings solvents may leak out. Please observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.
  • Page 25 Installing the Pump Flow Connections of the Binary Pump with Solvent Selection Valve 5 Using a piece of sanding paper connect the waste tubing to the purge valve and place it into your waste system. 6 If the binary pump is not part of a Agilent1100 System stack or placed on the bottom of a stack, connect the corrugated waste tube to the waste outlet of the pump leak handling system.
  • Page 26 Installing the Pump Flow Connections of the Binary Pump with Solvent Selection Valve Figure 6 Binary Pump with Solvent Selection Valve Bottle-head assembly Solvent cabinet Tube clip Static Mixer Channel B1 Channel A1 Purge valve Pump head Pump head channel A channel B Active inlet Active inlet valve A...
  • Page 27: Flow Connections Of The Binary Pump Without Solvent Selection Valve

    Installing the Pump Flow Connections of the Binary Pump without Solvent Selection Valve Flow Connections of the Binary Pump without Solvent Selection Valve WAR NI N G When opening capillary or tube fittings solvents may leak out. Please observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.
  • Page 28 Installing the Pump Flow Connections of the Binary Pump without Solvent Selection Valve 4 Connect the solvent tubes from the bottle head assemblies to the inlet adapters of the active inlet valves. Fix the tubes in the clips of solvent cabinet and binary pump.
  • Page 29 Installing the Pump Flow Connections of the Binary Pump without Solvent Selection Valve Figure 8 Flow Connection of Binary Pump Without Solvent Selection Valve Bottle-head assembly Solvent cabinet Tube clip Static Mixer Purge valve Pump head Pump head channel A channel B Adapter Active inlet...
  • Page 30: Priming And Purging The System

    Installing the Pump Priming and Purging the System Priming and Purging the System If a degasser is installed, it can be primed either by drawing solvent through the degasser with a syringe or by pumping with the pump. Priming the vacuum degasser or system with a syringe is recommended, when: •...
  • Page 31 Installing the Pump Priming and Purging the System 5 Push syringe adapter onto syringe. 6 Pull syringe plunger to draw at least 30 ml of solvent through degasser and tubing. 7 Replace the priming solvent with the new solvent of your choice. 8 Pull syringe plunger to draw at least 30 ml of solvent through degasser and tubing.
  • Page 32 Installing the Pump Priming and Purging the System 6 Close the purge valve and set the required composition and flow rate for your application. Table 4 Choice of Priming Solvents for Different Purposes Activity Solvent Comments After an installation Isopropanol Best solvent to flush air out of the system When switching between...
  • Page 33: Optimizing Performance

    Optimizing Performance How to optimize the binary pump to achieve best chromatographic results...
  • Page 34: Hints For Successful Use Of The Binary Pump

    Optimizing Performance Hints for Successful Use of the Binary Pump Hints for Successful Use of the Binary Pump • Place solvent cabinet with the solvent bottles always on top (or at a higher level) of the binary pump. • When using the binary pump without vacuum degasser, shortly degass your solvents (for example, water vacuum pump for 15 –...
  • Page 35 Optimizing Performance Hints for Successful Use of the Binary Pump of the seal. • After changing the plunger seals apply the seal wear-in procedure (see “Exchanging the Pump Seals and Seal Wear-in Procedure” on page 121). • Place the aqueous solvent on channel A and the organic solvent on channel B.
  • Page 36: Solvent Information

    Optimizing Performance Solvent Information Solvent Information Always filter solvents through 0.4 µm filters, small particles can permanently block the capillaries and valves. Avoid the use of the following steel-corrosive solvents: • Solutions of alkali halides and their respective acids (for example, lithium iodide, potassium chloride, and so on).
  • Page 37: Prevent Blocking Of Solvent Filters

    Optimizing Performance Prevent Blocking of Solvent Filters Prevent Blocking of Solvent Filters Contaminated solvents or algae growth in the solvent bottle will reduce the lifetime of the solvent filter and will influence the performance of the binary pump. This is especially true for aqueous solvents or phosphate buffers (pH 4 to 7).
  • Page 38 Optimizing Performance Prevent Blocking of Solvent Filters Cleaning the Solvent Filters • Remove the blocked solvent filter from the bottle-head assembly and place it in a beaker with concentrated nitric acid (35%) for one hour. • Thoroughly flush the filter with bidistilled water (remove all nitric acid, some capillary columns can be damaged by nitric acid).
  • Page 39: When To Use A Vacuum Degasser

    • binary pump was turned off for a length of time (for example, during night) and volatile solvent mixtures are used, or • solvents have been changed. For more information see the Reference Manual for the Agilent 1100 Series vacuum degasser.
  • Page 40: When To Use The Continuous Seal Wash Option

    Optimizing Performance When to Use the Continuous Seal Wash Option When to Use the Continuous Seal Wash Option Highly-concentrated buffer solutions will reduce the lifetime of the seals and plungers in your binary pump. The seal wash option allows to maintain the seal lifetime by flushing the back side of the seal with a wash solvent.
  • Page 41: When To Use Alternative Seals

    Optimizing Performance When to Use Alternative Seals When to Use Alternative Seals The standard seals for the binary pump can be used for most applications. However applications that use normal phase solvents (for example, hexane) are not suited for the standard seals and require a different seals when used for a longer time in the binary pump.
  • Page 42: When To Remove The Static Mixer

    Optimizing Performance When to Remove the Static Mixer When to Remove the Static Mixer The binary pump is equipped with a static mixer. The total delay volume of the pump is 600 – 900 µl. The mixer has a volume of 420 µl. The static mixer and both connecting capillaries can be replaced by a small capillary (G1312-67301) under the following conditions: •...
  • Page 43: How To Optimize The Compressibility Compensation Setting

    Optimizing Performance How to Optimize the Compressibility Compensation Setting How to Optimize the Compressibility Compensation Setting The compressibility compensation default settings are 50 × 10 /bar (best for most aqueous solutions) for pump head A and 115 × 10 /bar (to suit organic solvents) for pump head B.
  • Page 44 Optimizing Performance How to Optimize the Compressibility Compensation Setting Table 5 Solvent Compressibility Solvent (pure) Compressibility (10 /bar) Acetone Acetonitrile Benzene Carbon tetrachloride Chloroform Cyclohexane Ethanol Ethyl acetate Heptane Hexane Isobutanol Isopropanol Methanol 1-Propanol Toluene Water 6 Repeat step 1 through step 5 for the B channel of your binary pump.
  • Page 45 Troubleshooting and Test Functions The binary pump’s built-in troubleshooting and test functions...
  • Page 46: Troubleshooting And Test Functions

    Troubleshooting and Test Functions This chapter describes the instrument’s built in troubleshooting and test functions. Status Indicators The binary pump is provided with two status indicators which indicate the operational state (prerun, run, and error states) of the binary pump. The status indicators provide a quick visual check of the operation of the binary pump (see “Status Indicators”...
  • Page 47 Troubleshooting and Test Functions...
  • Page 48: Status Indicators

    Status Indicators Two status indicators are located on the front of the binary pump. The lower left one indicates the power supply status, the upper right one indicates the instrument status. Figure 9 Location of Status Indicators Status indicator Power supply indicator...
  • Page 49: Power Supply Indicator

    Troubleshooting and Test Functions Power Supply Indicator Power Supply Indicator The power supply indicator is integrated into the main power switch. When the indicator is illuminated (green) the power is ON. When the indicator is off, the module is turned OFF. Otherwhise check power connections, availability of power or check functioning of the power supply.
  • Page 50: Error Messages

    Error Messages Error messages are displayed in the user interface when an electronic, mechanical, or hydraulic (flow path) failure occurs which requires attention before the analysis can be continued (for example, repair, frit exchange or exchange of consumables required). In the event of such a failure, the red status indicator at the front of the module is switched on, and an entry is written into the instrument logbook.
  • Page 51: Timeout

    Troubleshooting and Test Functions Timeout Timeout The timeout threshold was exceeded. Probable Causes • The analysis was completed successfully, and the timeout function switched off the binary pump as requested. • A not-ready condition was present during a sequence or multiple-injection run for a period longer than the timeout threshold.
  • Page 52: Shutdown

    Suggested Actions Fix the leak in the external instrument before restarting the binary pump. Check external instruments for a shut-down condition. Check the vacuum degasser for an error condition. Refer to the Reference Manual for the Agilent 1100 Series vacuum degasser.
  • Page 53: Remote Timeout

    Troubleshooting and Test Functions Remote Timeout Remote Timeout A not-ready condition is still present on the remote input. When an analysis is started, the system expects all not-ready conditions (e.g. a not-ready condition during detector balance) to switch to run conditions within one minute of starting the analysis.
  • Page 54: Synchronization Lost

    Troubleshooting and Test Functions Synchronization Lost Synchronization Lost During an analysis, the internal synchronization or communication between one or more of the modules in the system has failed. The system processors continually monitor the system configuration. If one or more of the modules is no longer recognized as being connected to the system, the error message is generated.
  • Page 55: Leak

    Troubleshooting and Test Functions Leak Leak A leak was detected in the binary pump. The signals from the two temperature sensors (leak sensor and board-mounted temperature-compensation sensor) are used by the leak algorithm to determine whether a leak is present. When a leak occurs, the leak sensor is cooled by the solvent.
  • Page 56: Leak Sensor Open

    Troubleshooting and Test Functions Leak Sensor Open Leak Sensor Open The leak sensor in the binary pump has failed (open circuit). The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak-sensor current to change within defined limits.
  • Page 57: Leak Sensor Short

    Troubleshooting and Test Functions Leak Sensor Short Leak Sensor Short The leak sensor in the binary pump has failed (short circuit). The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak-sensor current to change within defined limits.
  • Page 58: Compensation Sensor Open

    Troubleshooting and Test Functions Compensation Sensor Open Compensation Sensor Open The ambient-compensation sensor (NTC) on the HPM board in the binary pump has failed (open circuit). The resistance across the temperature compensation sensor (NTC) on the HPM board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes.
  • Page 59: Compensation Sensor Short

    Troubleshooting and Test Functions Compensation Sensor Short Compensation Sensor Short The ambient-compensation sensor (NTC) on the HPM board in the binary pump has failed (short circuit). The resistance across the temperature compensation sensor (NTC) on the HPM board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes.
  • Page 60: Fan Failed

    Troubleshooting and Test Functions Fan Failed Fan Failed The cooling fan in the binary pump has failed. The hall sensor on the fan shaft is used by the HPM board to monitor the fan speed. If the fan speed falls below 2 revolutions/second for longer than 5 seconds, the error message is generated.
  • Page 61: Open Cover

    Troubleshooting and Test Functions Open Cover Open Cover The top foam has been removed. The sensor on the HPM board detects when the top foam is in place. If the foam is removed, the fan is switched off, and the error message is generated. Probable Causes •...
  • Page 62: Restart Without Cover

    Troubleshooting and Test Functions Restart Without Cover Restart Without Cover The binary pump was restarted with the top cover and foam open. The sensor on the HPM board detects when the top foam is in place. If the binary pump is restarted with the foam removed, the binary pump switches off within 30 s, and the error message is generated.
  • Page 63: Zero Solvent Counter

    Troubleshooting and Test Functions Zero Solvent Counter Zero Solvent Counter Pump firmware version A.02.32 and higher allow to set solvent bottle fillings at the ChemStation (revision 5.xx and higher). If the volume level in the bottle falls below the specified value the error message appears when the feature is configured accordingly.
  • Page 64: Pressure Above Upper Limit

    Troubleshooting and Test Functions Pressure Above Upper Limit Pressure Above Upper Limit The system pressure has exceeded the upper pressure limit. Probable Causes • Upper pressure limit set too low. • Blockage in the flowpath (after the damper). • Defective damper. •...
  • Page 65: Pressure Below Lower Limit

    Troubleshooting and Test Functions Pressure Below Lower Limit Pressure Below Lower Limit The system pressure has fallen below the lower pressure limit. Probable Causes • Lower pressure limit set too high. • Air bubbles in the mobile phase. • Leak. •...
  • Page 66: Pressure Signal Missing

    Troubleshooting and Test Functions Pressure Signal Missing Pressure Signal Missing The pressure signal from the damper is missing. The pressure signal from the damper must be within a specific voltage range. If the pressure signal is missing, the processor detects a voltage of approximately -120mV across the damper connector.
  • Page 67: Valve Failed

    Troubleshooting and Test Functions Valve Failed Valve Failed Valve 0 Failed: valve A1 Valve 1 Failed: valve A2 Valve 2 Failed: valve B2 Valve 3 Failed: valve B1 One of the solvent selection valves in the binary pump failed to switch correctly.
  • Page 68: Missing Pressure Reading

    Troubleshooting and Test Functions Missing Pressure Reading Missing Pressure Reading The pressure readings read by the pump ADC (analog-digital converter) are missing. The ADC reads the pressure readings from the damper every 1ms. If the readings are missing for longer than 10 seconds, the error message is generated.
  • Page 69: Pump Configuration

    Troubleshooting and Test Functions Pump Configuration Pump Configuration At switch-on, the pump has recognized a new pump configuration. The binary pump is assigned its configuration at the factory. If the active-inlet valve and pump encoder of channel B are disconnected, and the binary pump is rebooted, the error message is generated.
  • Page 70: Valve Fuse

    Troubleshooting and Test Functions Valve Fuse Valve Fuse Valve Fuse 0: Channels A1 and A2 Valve Fuse 1: Channels B1 and B2 One of the solvent-selection valves in the binary pump has drawn excessive current causing the selection-valve electronic fuse to open. Probable Causes •...
  • Page 71: Inlet-Valve Fuse

    Troubleshooting and Test Functions Inlet-Valve Fuse Inlet-Valve Fuse Inlet-Valve Fuse 0: Pump channel A Inlet-Valve Fuse 1: Pump channel B One of the active-inlet valves in the binary pump has drawn excessive current causing the inlet-valve electronic fuse to open. Probable Causes •...
  • Page 72: Temperature Out Of Range

    Troubleshooting and Test Functions Temperature Out of Range Temperature Out of Range Temperature Out of Range 0: Pump channel A Temperature Out of Range 1: Pump channel B One of the temperature sensor readings in the motor-drive circuit are out of range.
  • Page 73: Temperature Limit Exceeded

    Troubleshooting and Test Functions Temperature Limit Exceeded Temperature Limit Exceeded Temperature Limit Exceeded 0: Pump channel A Temperature Limit Exceeded 1: Pump channel B The temperature of one of the motor-drive circuits is too high. The processor continually monitors the temperature of the drive circuits on the HPM board.
  • Page 74: Motor-Drive Power

    Troubleshooting and Test Functions Motor-Drive Power Motor-Drive Power Motor-Drive Power: Pump channel A B: Motor-Drive Power: Pump channel B The current drawn by the pump motor exceeded the maximum limit. Blockages in the flowpath are usually detected by the pressure sensor in the damper, which result in the binary pump switching off when the upper pressure limit is exceeded.
  • Page 75: Encoder Missing

    Troubleshooting and Test Functions Encoder Missing Encoder Missing Encoder Missing: Pump channel A B: Encoder Missing: Pump channel B The optical encoder on the pump motor in the binary pump is missing or defective. The processor checks the presence of the pump encoder connector every 2 seconds.
  • Page 76: Inlet-Valve Missing

    Troubleshooting and Test Functions Inlet-Valve Missing Inlet-Valve Missing Inlet-Valve Missing: Pump channel A B: Inlet-Valve Missing: Pump channel B The active-inlet valve in the binary pump is missing or defective. The processor checks the presence of the active-inlet valve connector every 2 seconds.
  • Page 77: Servo Restart Failed

    Troubleshooting and Test Functions Servo Restart Failed Servo Restart Failed Servo Restart Failed: Pump channel A B: Servo Restart Failed: Pump channel B The pump motor in the binary pump was unable to move into the correct position for restarting. When the binary pump is switched on, the first step is to switch on the C phase of the variable reluctance motor.
  • Page 78: Pump Head Missing

    Troubleshooting and Test Functions Pump Head Missing Pump Head Missing Pump Head Missing: Pump channel A B: Pump Head Missing: Pump channel B The pump-head end stop in the binary pump was not found. When the binary pump restarts, the metering drive moves forward to the mechanical end stop.
  • Page 79: Index Limit

    Troubleshooting and Test Functions Index Limit Index Limit Index Limit: Pump channel A B: Index Limit: Pump channel B The time required by the plunger to reach the encoder index position was too short (binary pump). During initialization, the first plunger is moved to the mechanical stop. After reaching the mechanical stop, the plunger reverses direction until the encoder index position is reached.
  • Page 80: Index Adjustment

    Troubleshooting and Test Functions Index Adjustment Index Adjustment Index Adjustment : Pump channel A B: Index Adjustment: Pump channel B The encoder index position in the binary pump is out of adjustment. During initialization, the first plunger is moved to the mechanical stop. After reaching the mechanical stop, the plunger reverses direction until the encoder index position is reached.
  • Page 81: Index Missing

    Troubleshooting and Test Functions Index Missing Index Missing Index Missing: Pump channel A B: Index Missing: Pump channel B The encoder index position in the binary pump was not found during initialization. During initialization, the first plunger is moved to the mechanical stop. After reaching the mechanical stop, the plunger reverses direction until the encoder index position is reached.
  • Page 82: Stroke Length

    Troubleshooting and Test Functions Stroke Length Stroke Length Stroke Length: Pump channel A B: Stroke Length: Pump channel B The distance between the lower plunger position and the upper mechanical stop is out of limits (binary pump). During initialization, the binary pump monitors the drive current. If the plunger reaches the upper mechanical stop position before expected, the motor current increases as the binary pump attempts to drive the plunger beyond the mechanical stop.
  • Page 83: Initialization Failed

    Troubleshooting and Test Functions Initialization Failed Initialization Failed Initialization Failed: Pump channel A B: Initialization Failed: Pump channel B The binary pump failed to initialize successfully within the maximum time window. A maximum time is assigned for the complete pump-initialization cycle. If the time is exceeded before initialization is complete, the error message is generated.
  • Page 84: Wait Timeout

    Wait Timeout When running certain tests in the diagnostics mode or other special applications, the pump must wait for the plungers to reach a specific position, or must wait for a certain pressure or flow to be reached. Each action or state must be completed within the timeout period, otherwise the error message is generated.
  • Page 85: Pressure Test

    Troubleshooting and Test Functions Pressure Test Pressure Test Description The pressure test is a quick, built-in test designed to demonstrate the pressure-tightness of the system. The test should be used when problems with small leaks are suspected, or after maintenance of flow-path components (e.g., pump seals, injection seal) to prove pressure tightness up to 400 bar.
  • Page 86 Troubleshooting and Test Functions Pressure Test Figure 10 Typical Pressure-Test Pressure Profile with IPA Pressure Step 2 Step 1 Time [minutes] Step 1 The test begins with the initialization of both pumpheads. After initialization, plungers A1 and B1 are both at the top of their stroke. Next, pump A begins pumping solvent with a flow rate of 510 µl/min and stroke of 100 µl.
  • Page 87: Running The Pressure Test

    Troubleshooting and Test Functions Running the Pressure Test Running the Pressure Test Tools required Wrench 1/4 inch. Parts and materials Blank nut, 01080-83202 required Isopropanol, 500 ml N O T E Make absolutely sure that all parts of the flow path that are part of the test are very thoroughly flushed with IPA before starting to pressurize the system! Any trace of other solvents or the smallest air bubble inside the flow path definitely will cause the test to fail!
  • Page 88 Troubleshooting and Test Functions Running the Pressure Test 5 For binary pumps without degasser set flow for channel A or A2 to 5 ml/min and flush pump head A for about 2 minutes. 6 Set flow to 0 ml/min. Leave the purge valve open. 7 Connect the signal cable to the analog output at the rear of the binary pump (only if an integrator is used).
  • Page 89: Evaluating The Results

    Troubleshooting and Test Functions Evaluating the Results Evaluating the Results The sum of all leaks between the pump and the blank nut will be indicated by a pressure drop of >2 bar/minute at the plateau. Note that small leaks may cause the test to fail, but solvent may not be seen leaking from a module.
  • Page 90 Troubleshooting and Test Functions Evaluating the Results Potential Causes of Pressure Test Failure After isolating and fixing the cause of the leak, repeat the pressure test to confirm the system is pressure tight. Potential Cause (Pump) Corrective Action Purge valve open. Close the purge valve.
  • Page 91: Leak Test

    Troubleshooting and Test Functions Leak Test Leak Test The leak test is a built-in troubleshooting test designed to demonstrate the leak-tightness of the binary pump. The test should be used when problems with the binary pump are suspected. The test involves monitoring the pressure profile as the binary pump runs through a predefined pumping sequence.
  • Page 92: Description

    Description The pump outlet is blocked with a blank nut, and then the test is run using isopropyl alcohol (IPA), while monitoring the pressure profile (using an integrator on the analog output, or in the plot screen in the Control Module or the ChemStation).
  • Page 93 Troubleshooting and Test Functions Description should be linear. Large leaks or defects will be characterized by an unstable, non-linear slope. Ramp 3 Just before the start of the first plateau, plunger A2 delivers with a flow rate of 50 µl/min for approximately 8 seconds. The system pressure should be 240 bar or higher.
  • Page 94: Running The Leak Test

    Troubleshooting and Test Functions Running the Leak Test Running the Leak Test Tools required Wrench 1/4 inch Parts and materials Restriction Capillary ,G1313-87305 required Blank nut, 01080-83202 Isopropanol, 500ml N O T E Make absolutely sure that all parts of the flow path that are part of the test are very thoroughly flushed with IPA before starting to pressurize the system! Any trace of other solvents or the smallest air bubble inside the flow path definitely will cause the test to fail!
  • Page 95 Troubleshooting and Test Functions Running the Leak Test Binary Pumps without Degasser 5 Set Flow to 5 ml/min for channel A or A2 and flush pump head A for about 2 minutes. 6 Set Flow to 5 ml/min for channel B or B2 and flush pump head B for about 2 minutes.
  • Page 96: Evaluating The Results

    Troubleshooting and Test Functions Evaluating the Results Evaluating the Results Defective or leaky components in the pump head lead to changes in the leak-test pressure plot. Typical failure modes are described below. Figure 12 Leak Test Pressure Plot Pressure [bar] Plateau 4 Plateau 3 Plateau 2...
  • Page 97 Troubleshooting and Test Functions Evaluating the Results N O T E The pressure plots shown below are examples only. The plots may vary depending on the type and degree of leakage. No pressure increase or minimum pressure of plateau 1 not reached Potential Cause Corrective Action Pump not running.
  • Page 98 Troubleshooting and Test Functions Evaluating the Results All plateaus negative Potential Cause Corrective Action Loose or leaky fittings. Ensure all fittings are tight, or exchange capillary. Loose purge valve. Tighten the purge valve (14mm wrench). Leaky mixer (if installed). Tighten the mixer fittings and nuts. Contaminated purge valve.
  • Page 99 Troubleshooting and Test Functions Evaluating the Results First plateau negative or unstable, and at least one other plateau positive. Potential Cause Corrective Action Leaking outlet valve in channel A. Clean the outlet valve in channel A. Ensure the sieve in the outlet valves are installed correctly.
  • Page 100 Troubleshooting and Test Functions Evaluating the Results Second plateau negative or unstable, and at least one other plateau positive Potential Cause Corrective Action Leaking outlet valve in channel B. Clean the outlet valve in channel B. Ensure the sieve in the outlet valves are installed correctly.
  • Page 101 Troubleshooting and Test Functions Evaluating the Results Third plateau negative or unstable and at least one other plateau positive Potential Cause Corrective Action Air in channel A or new seals not yet seated. Flush channel A thoroughly with isopropanol under pressure (use restriction capillary). Loose active inlet valve in channel A.
  • Page 102 Troubleshooting and Test Functions Evaluating the Results Fourth plateau negative or unstable and at least one other plateau positive Potential Cause Corrective Action Air in pump chamber of channel B or seals Flush channel B thoroughly with isopropanol not yet seated. under pressure (restriction capillary).
  • Page 103: Repairing The Pump

    Repairing the Pump Instructions on simple, routine repair procedures as well as more extensive repairs requiring exchange of internal parts...
  • Page 104 Repairing the Pump Simple Repairs The binary pump is designed for easy repair. The most frequent repairs such as plunger seal change and purge valve frit change can be done from the front of the binary pump with the binary pump in place in the system stack. These repairs are described in “Simple Repair Procedures”...
  • Page 105: Cleaning The Binary Pump

    Repairing the Pump Cleaning the Binary Pump The binary pump case should be kept clean. Cleaning should be done with a soft cloth slightly dampened with water or a solution of water and a mild detergent. Do not use an excessively damp cloth that liquid can drip into the binary pump.
  • Page 106: Using The Esd Strap

    Repairing the Pump Using the ESD Strap Using the ESD Strap Electronic boards are sensitive to electrostatic discharge (ESD). In order to prevent damage, always use an ESD strap supplied in the standard accessory kit (see “Accessory Kit G1311-68705” on page 186) when handling electronic boards and components.
  • Page 107: Overview

    Repairing the Pump Overview Overview Figure 14 shows the main assemblies of the binary pump. The pump heads and its parts do require normal maintenance (for example, seal exchange) and can be accessed from the front (simple repairs). Replacing internal parts will require to remove the module from its stack and to open the top cover.
  • Page 108: Simple Repair Procedures

    Simple Repair Procedures The procedures described in this section can be done with the binary pump in place in the system stack. Table 6 Simple Repair Procedures Procedure Typical Frequency Notes “Exchanging the Active Inlet Valve Cartridge If internally leaking Pressure ripple unstable, run leak test for or the Active Inlet Valve”...
  • Page 109: Exchanging The Active Inlet Valve Cartridge Or The Active Inlet Valve

    Repairing the Pump Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve When required If internally leaking (backflow) Tools required Wrench 14 mm Material Active inlet valve G1312-60010 Valve Cartridge 5062-8562 Removing the Active Inlet Valve 1 Unplug the active inlet valve cable from the connector.
  • Page 110 Repairing the Pump Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve Exchanging the Valve Cartrigde 1 Using a pair of tweezers remove the valve cartridge from the actuator assembly. 2 Before inserting the new valve cartridge clean the area in the actuator assembly.
  • Page 111 Repairing the Pump Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve Figure 16 Exchanging the Active Inlet Valve Plug Solvent inlet tube Active inlet valve...
  • Page 112: Exchanging The Outlet Ball Valve Sieve Or The Complete Valve

    Repairing the Pump Exchanging the Outlet Ball Valve Sieve or the Complete Valve Exchanging the Outlet Ball Valve Sieve or the Complete Valve When required Sieve — whenever the pump seals will be exchanged Valve — if internally leaking Tools required Wrench 1/4 inch Wrench 14 mm Parts required...
  • Page 113 Repairing the Pump Exchanging the Outlet Ball Valve Sieve or the Complete Valve Figure 17 Outlet Ball Valve Parts Valve body Sieve Gold seal Plastic cap 7 Reinstall the outlet ball valve and tighten the valve. 8 Reconnect the valve capillary. Figure 18 Exchanging the Outlet Ball Valve Valve capillary...
  • Page 114: Exchanging The Purge Valve Frit Or The Purge Valve

    Repairing the Pump Exchanging the Purge Valve Frit or the Purge Valve Exchanging the Purge Valve Frit or the Purge Valve When required Frit - when plunger seals are exchanged or when contaminated or blocked (pressure drop of > 10 bar across the frit at a flow rate of 5 ml/min of H 0 with purge valve opened) Purge valve - if internally leaking Tools required...
  • Page 115 Repairing the Pump Exchanging the Purge Valve Frit or the Purge Valve Figure 19 Purge Valve Parts Valve body PTFE frit Gold seal Plastic cap 6 Place a new frit into the purge valve with the orientation of the frit as shown above.
  • Page 116 Repairing the Pump Exchanging the Purge Valve Frit or the Purge Valve Figure 20 Exchanging the Purge Valve Frit Purge valve holder Pump outlet capillary Purge valve Waste capillary...
  • Page 117: Exchanging The Solvent Selection Valve

    Repairing the Pump Exchanging the Solvent Selection Valve Exchanging the Solvent Selection Valve When required If internally leaking ( croossflow between the ports), or if one of the channels is blocked Tools required Screwdriver Pozidriv #1 Parts required Solvent selection valve (PN gives half of a complete solvent selection block) G1312-60000 1 Disconnect the solvent tubes and the active inlet valve connection tubes from the solvent selection valves.
  • Page 118 Repairing the Pump Exchanging the Solvent Selection Valve Figure 21 Exchanging the solvent selection valve Solvent tubes Solvent selection valve Connecting tubes...
  • Page 119: Removing And Disassembling The Pump Head Assembly

    Repairing the Pump Removing and Disassembling the Pump Head Assembly Removing and Disassembling the Pump Head Assembly WAR NI N G Never start the pump when the pump head is removed. This may damage the pump drive. N O T E Both pump head assemblies use the same internal components.
  • Page 120 Repairing the Pump Removing and Disassembling the Pump Head Assembly 2 Using a 3-mm hexagonal key loosen the purge 3 Using a 4-mm hexagonal key stepwise loosen valve holder and lift it up. and remove the two pump head screws and remove the pump head from the pump drive.
  • Page 121: Exchanging The Pump Seals And Seal Wear-In Procedure

    Repairing the Pump Exchanging the Pump Seals and Seal Wear-in Procedure Exchanging the Pump Seals and Seal Wear-in Procedure 1 Disassemble the pump head assembly of the When required: leaky pump head (see “Removing and Seals leaking, if indicated by the results of Disassembling the Pump Head Assembly”...
  • Page 122 Repairing the Pump Exchanging the Pump Seals and Seal Wear-in Procedure 2 Using one of the plungers carefully remove 3 Clean the pump chambers with lint free cloth. the seal from the pump head (be careful, not Ensure all particulate matter is removed. Best to break the plunger).
  • Page 123 Repairing the Pump Exchanging the Pump Seals and Seal Wear-in Procedure Seal Wear-in Procedure N O T E This procedure is required for standard seals only (5063-6589), but it will definetely damage the normal phase application seals (0905-1420). 1 Place a bottle with 100 ml of Isopropanol in the solvent cabinet and place the tubing (including bottle head assembly) of the pump head that is supposed to be worn-in into the bottle.
  • Page 124: Exchanging The Plungers

    Repairing the Pump Exchanging the Plungers Exchanging the Plungers 1 Disassemble the pump head assembly (see When required: ““Removing and Disassembling the Pump When scratched Head Assembly” on page 119). Tools required: 3-mm hexagonal key 4-mm hexagonal key Parts: Plunger 5063-6586 2 Check the plunger surface and remove any 3 Reassemble the pump head assembly (see deposits or layers.
  • Page 125: Installing The Continuous Seal Wash Option

    Repairing the Pump Installing the Continuous Seal Wash Option Installing the Continuous Seal Wash Option 1 Disassemble the pump head assembly (see Tools required: “Removing and Disassembling the Pump 4-mm hexagonal key Head Assembly” on page 119). 3-mm hexagonal key Parts: Seal wash kit (01018-68722) 2 Remove the support rings from the plunger...
  • Page 126 Repairing the Pump Installing the Continuous Seal Wash Option 4 Install the seal wash support ring assembly 5 If necessary replace the pump seals (see from the seal wash option kit into the plunger page 121 ). Reassemble the pump head housing.
  • Page 127: Exchanging The Wash Seals

    Repairing the Pump Exchanging the Wash Seals Exchanging the Wash Seals 1 Disassemble the pump head assembly (see When required: “Removing and Disassembling the Pump If leaking Head Assembly” on page 119). Tools required: 4-mm hexagonal key 3-mm hexagonal key Insert tool Small flat-head screwdriver Parts:...
  • Page 128 Repairing the Pump Exchanging the Wash Seals 4 Using the insert tool press the seal (spring 5 Reassemble the pump head assembly (see pointing upwards) into the recess of the “Reassembling the Pump Head Assembly” on support ring. Place a seal wash gasket in the page 129).
  • Page 129: Reassembling The Pump Head Assembly

    Repairing the Pump Reassembling the Pump Head Assembly Reassembling the Pump Head Assembly 1 Place the support rings on the plunger Tools required: housing (plungers not installed) and snap the 3-mm hexagonal key pump head and plunger housing together. 4-mm hexagonal key PTFE lubricant (79841-65501) Pump head Support ring...
  • Page 130 Repairing the Pump Reassembling the Pump Head Assembly 4 Slide the pump head assembly onto the pump 5 Reconnect the capillaries, tubing and the drive. Apply a small amount of pump head active inlet valve cable to the connector. grease to the pumphead screws and the balls of the spindle drive.
  • Page 131: Exchanging The Optional Interface Board

    Repairing the Pump Exchanging the Optional Interface Board Exchanging the Optional Interface Board CA UT IO N The interface board is sensitive to electrostatic discharge. Always use the ESD kit when handling electronic boards. When required Board defective Part required BCD (Interface) board, see “Optional Interface Boards”...
  • Page 132: Exchanging Internal Parts

    Repairing the Pump Exchanging Internal Parts Exchanging Internal Parts WAR NI N G The following procedures require opening the main cover of the binary pump. Always ensure the binary pump is disconnected from the line power when the main cover is removed. The security lever at the power input socket prevents that the pump cover is taken off when line power is still connected.
  • Page 133: Removing The Top Cover And Foam

    Repairing the Pump Removing the Top Cover and Foam Removing the Top Cover and Foam Tools required Screwdriver Pozidriv #1 Preparations for this Switch off binary pump at the main power switch. procedure Disconnect the solvent inlet tubes from the solvent selection valve or the adapter at the active inlet valve.
  • Page 134 Repairing the Pump Removing the Top Cover and Foam 3 Lift the clips on both sides of the top cover 4 Unscrew the screws on the top plate and (1). Remove the top cover (2). remove the plate by lifting its back first and then sliding to the front.
  • Page 135 Repairing the Pump Removing the Top Cover and Foam Do not connect a power plug to the binary pump 7 Position of the safety switch on the main module after removing the top covers. board. A safety light switch on the main board will prevent operation when the covers are removed.
  • Page 136: Exchanging The High Pressure Pump Main Board (Hpm Board)

    Repairing the Pump Exchanging the High Pressure Pump Main Board (HPM Board) Exchanging the High Pressure Pump Main Board (HPM Board) When required Board defective Tools required Wrench 14 mm Wrench 7 mm Wrench 5 mm Parts required HPM board, G1312-66520, exchange part number G1312-69520 1 Turn off the pump, disconnect all cables and remove the pump from the stack.
  • Page 137 Repairing the Pump Exchanging the High Pressure Pump Main Board (HPM Board) Figure 23 Board Layout of the HPM Board 4 Remove the connector screws from the GPIB, Remote and from the analog pressure output connector. 5 Remove the board. Place it on an ESD kit. WAR NI N G The RFI spring plate sitting on the board connectors is very sharp! Be careful, not to cut yourself, when removing it from the old board and...
  • Page 138 Exchanging the High Pressure Pump Main Board (HPM Board) 8 On the new board check the switch setting of address switch S1, see “Default Addresses for Agilent 1100 Series Modules” on page 231. N O T E An incorrect switch setting (e.g., TEST/BOOT) may cause the pump to turn into a basic mode (yellow or red flashing status light).
  • Page 139 Repairing the Pump Exchanging the High Pressure Pump Main Board (HPM Board) Entering the Type Command N O T E After the installation of a new mainboard the TYPE (binary) of the module is normally automatically detected. The specific TYPE tells the pump how to configure itself during turn on.
  • Page 140 Repairing the Pump Exchanging the High Pressure Pump Main Board (HPM Board) 9 Turn off the module, then turn it on again. Turn on should be normal. In the Records screen, the product# column should indicate the binary pump. If a ChemStation is also connected, re-boot it now.
  • Page 141 Repairing the Pump Exchanging the High Pressure Pump Main Board (HPM Board) 5 Letters and numbers are created using the up and down arrows. Into the box labeled Serial#, enter the 10-character serial number for the binary pump. When the 10-character serial number is entered, press Enter to highlight the complete serial number.
  • Page 142 Repairing the Pump Exchanging the High Pressure Pump Main Board (HPM Board) Replacing the Binary Pump’s Firmware The installation of new firmware is required • if a new version solves problems of the currently installed version. • if the version of firmware on the new main board (HPM) after an exchange of the board is older than the one previously installed.
  • Page 143: Exchanging The Damper

    Repairing the Pump Exchanging the Damper Exchanging the Damper When required: No pressure output or when leaking Tools required: Screwdriver Pozidriv #1 Wrench 1/4 inch Parts required: Damper, PN 79835-60005 N O T E Do not totally remove the Z-panel when the damper is in its place. The heavy damper may fall down..
  • Page 144 Repairing the Pump Exchanging the Damper 3 Loosen the screws of the Z-panel, fold it 4 Replace the Z-panel and the new damper and forward and remove the damper. fix it with the two screws. Reconnect the damper connector (J24) at the main board. Screw Damper Screw...
  • Page 145: Exchanging The Fan

    Repairing the Pump Exchanging the Fan Exchanging the Fan When required: Fan not running Tools required: Screwdriver Pozidriv #1 Wrench 1/4 inch Parts required: Fan, PN 3160-1017 1 Remove the front cover, top cover and top 2 Disconnect the capillaries from pump head B. foam section (see “Removing the Top Cover Disconnect both active inlet valve cables and and Foam”...
  • Page 146 Repairing the Pump Exchanging the Fan 3 Loosen the screws of the Z-panel, fold it 4 Disconnect the cables of pump B (J12, J27) forward and remove the damper. Disconnect from the main board and lift out the complete the two flat cables of the active inlet valve assembly.
  • Page 147 Repairing the Pump Exchanging the Fan 7 Place pump B back into its recess and 8 Replace the Z-panel and the damper and fix it connect to the main board. with the two screws. Reconnect the damper connector (J24) and the AIV cables at the main board (J19, J28).
  • Page 148: Exchanging A Pump Drive

    Repairing the Pump Exchanging a Pump Drive Exchanging a Pump Drive WAR NI N G Never start the pump when the pump head is removed. This may damage the pump drive. The binary pump has two pump-drive assemblies. The A drive is located on the left side and the B drive on the right side.
  • Page 149 Repairing the Pump Exchanging a Pump Drive 2 Disconnect all capillaries and the waste 3 Using a 4-mm hexagonal key, stepwise loosen tubing from the pump head and damper and and remove the two pump head screws and disconnect the active inlet valve connectors remove the pump head from the pump drive.
  • Page 150 Repairing the Pump Exchanging a Pump Drive 6 Place the new pump drive into the recess in 7 Replace the damper and the Z-panel and the foam part and connect the cables to the secure it with the two screws. Reconnect the connectors on the main board (J16, J17 or cables to the main board (J19, J28).
  • Page 151: Exchanging The Power Supply

    Repairing the Pump Exchanging the Power Supply Exchanging the Power Supply When required: If defective Tools required: Screwdriver Pozidriv #1 Wrench 1/4 inch Wrench 14 mm Wrench 7 mm Wrench 5 mm Part required: Power supply 0950-2528 1 Disconnect all cables and capillaries, remove 2 Disconnect all capillaries, tubes and the pump from stack, remove the front cover, connectors from the two pump heads and the...
  • Page 152 Repairing the Pump Exchanging the Power Supply 3 Disconnect the connectors of the active inlet 4 Disconnect both pumps from the main board valve and the damper from the main board (J16, J17, J12, J28) and lift them out of the (J19, J24, J28), loosen the Z-panel screws and foam.
  • Page 153 Repairing the Pump Exchanging the Power Supply 7 Push the leak sensor cable through the recess 8 Loosen and remove the power-supply screws of the solvent selection valve and lift out the at the rear panel. bottom foam. Leak sensor cable Screws 9 Unclip the power supply light pipe from the...
  • Page 154 Repairing the Pump Exchanging the Power Supply 11 Place the new power supply into the 12 Place the coupler onto the switch in the instrument and fix it with the two screws at power supply and clip the light pipe back the rear panel.
  • Page 155 Repairing the Pump Exchanging the Power Supply 15 Re-install solvent selection valve, if present 16 Replace the two pump units and connect to (see “Exchanging the Solvent Selection the main board (J16, J17, J12, J27). Valve” on page 117). 17 Place Z-panel into its position and replace the 18 Reconnect all capillaries, tubes and cables to damper.
  • Page 156: Exchanging The Leak Sensor

    Repairing the Pump Exchanging the Leak Sensor Exchanging the Leak Sensor When required Leak messages without leak in the funnel Tools required Screwdriver Pozidriv #1 Wrench 1/4 inch Wrench 14 mm Wrench 7 mm Wrench 5 mm Parts required Leak sensor, 5061-3356 1 Remove the top covers and foam, see“Removing the Top Cover and Foam”...
  • Page 157 Repairing the Pump Exchanging the Leak Sensor Figure 25 Exchanging the Leak Sensor To main board J23 Leak sensor Leak pan...
  • Page 158: Exchanging Status Light Pipe

    Repairing the Pump Exchanging Status Light Pipe Exchanging Status Light Pipe When required: If part is broken Tools required: Screwdriver Pozidriv #1 Part required: Status light pipe 5041-8384 Preparation for this procedure: Remove the front cover and top cover, see “Removing the Top Cover and Foam” on page 133. 1 The status light pipe is clipped into the top 2 Replace the top cover, see “Replacing the Top cover.
  • Page 159: Assembling The Main Cover

    Repairing the Pump Assembling the Main Cover Assembling the Main Cover When required If cover is broken Tools required None Parts required Cover kit G1312-68703 (includes base, top, left and right) N O T E The cover kit contains all parts, but it is not assembled. WAR NI N G In case you insert the left or right side in the opposite position, you may not be able to remove the side from the top part.
  • Page 160: Replacing The Top Cover And Foam

    Repairing the Pump Replacing the Top Cover and Foam Replacing the Top Cover and Foam Tools required Screwdriver Pozidriv #1 Preparations for this Make sure that after your repair all assemblies, cables, capillaries and connectors are procedure located in its correct place. 1 Place the damper cable in a curve to the right 2 Make sure that the foam is installed correctly side towards the active inlet valve cable of...
  • Page 161 Repairing the Pump Replacing the Top Cover and Foam 3 Location of foam in the light switch. 4 Replace the optional interface board or the board cover plate. Light switch 5 Replace the metal cover (slide the metal tabs 6 Replace the top cover. into place (1) underneath the Z-Panel in the front, then lower the back of the metal plate (2)) and fix the two holding screws.
  • Page 162 Repairing the Pump Replacing the Top Cover and Foam 7 Ensure clips are seated correctly, move lever 8 Replace the leak funnel with the waste tube. back. Locate the lower end of the waste tube in the holder of the leak pan. Leak funnel Lever Leak pan...
  • Page 163: Parts And Materials

    Parts and Materials Detailed illustrations and lists for identification of parts and materials...
  • Page 164 Parts and Materials Figure 26 Overview of Main Assemblies (Front View)
  • Page 165 Parts and Materials Table 7 Repair Parts — Pump Housing and Main Assemblies (Front View) Item Description Part Number Pump head, see page 179 G1311-60004 Pump drive assembly G1311-60001 Exchange assembly — pump drive G1311-69001 Cable assembly — AIV to main board G1311-61601 High-pressure pump main board (HPM) G1312-66520...
  • Page 166 Parts and Materials Figure 27 Overview of Main Assemblies (Rear View) Table 8 Repair Parts — Pump Housing and Main Assemblies (Rear View) Item Description Part Number Nut M14 — analog output 2940-0256 Washer — analog output 2190-0699 Standoff — remote connector 1251-7788 Standoff —...
  • Page 167: Control Module(B-Version)

    Parts and Materials Control Module(B-Version) Control Module(B-Version) Table 9 Control Module Item Description Part Number Control module, replacement part including cable G1323-67001 Plastic housing kit, includes front, back and clamp 5062-8583 CAN cable, Agilent 1100 module to control module G1323-81600 Figure 28 Control Module...
  • Page 168: Solvent Cabinet

    Parts and Materials Solvent Cabinet Solvent Cabinet Table 10 Solvent Cabinet Parts Item Description Part Number Solvent cabinet, including all plastic parts 5062-8581 Name plate, Agilent 1100 5042-1312 Front panel, solvent cabinet 5062-8580 Leak pan, solvent cabinet 5042-1307 Bottle amber 9301-1450 Bottle transparent 9301-1420...
  • Page 169: Bottle-Head Assembly

    Parts and Materials Bottle-Head Assembly Bottle-Head Assembly Table 11 Bottle-Head Assembly Parts Item Description Part Number Complete assembly G1311-60003 Ferrules with lock ring 5063-6598 (10x) Tube screw 5063-6599 (10x) Wire marker No part number Solvent tubing, 5 m 5062-2483 Frit adapter (pack of 4) 5062-8517 Solvent inlet filter 5042-1347...
  • Page 170: Hydraulic Path With Solvent Selection Valve

    Parts and Materials Hydraulic Path with Solvent Selection Valve Hydraulic Path with Solvent Selection Valve Table 12 Hydraulic Path with Solvent Selection Valve Item Description Part Number Capillary, mixer to purge valve G1312-67308 Capillary, outlet ball valve to piston 2 G1312-67300 Mixer G1312-87330...
  • Page 171 Parts and Materials Hydraulic Path with Solvent Selection Valve Figure 31 Hydraulic Path with Solvent Selection Valve...
  • Page 172: Hydraulic Path Without Solvent Selection Valve

    Parts and Materials Hydraulic Path without Solvent Selection Valve Hydraulic Path without Solvent Selection Valve Table 13 Hydraulic Path without Solvent Selection Valve Item Description Part Number Capillary, mixer to purge valve G1312-67308 Capillary, outlet ball valve to piston 2 G1312-67300 Mixer G1312-87330...
  • Page 173 Parts and Materials Hydraulic Path without Solvent Selection Valve Figure 32 Hydraulic Path without Solvent Selection Valve...
  • Page 174: Cover Parts

    Parts and Materials Cover Parts Cover Parts Table 14 Cover Parts Item Description Part Number Plastic cover kit (includes top, base and both sides) G1312-68703 Front plate G1312-60011 Logo plate, Agilent 1100 5042-1312 Figure 33 Cover Parts 1 (top) 1 (side) 1 (base)
  • Page 175: Sheet Metal Kit

    Parts and Materials Sheet Metal Kit Sheet Metal Kit Table 15 Sheet Metal Kit Item Description Part Number Sheet metal kit, includes top, base and Z-panel G1312-68701 Screw, for cover and Z-Panel 5022-2112 Board cover 5001-3772 Included in item 1 Figure 34 Sheet Metal Kit Z-panel...
  • Page 176: Foam Parts

    Parts and Materials Foam Parts Foam Parts Table 16 Foam Parts Item Description Part Number Foam kit, including upper and lower foam G1312-68702 Bushing, for pump drive 1520-0404 Board guide 5041-8395 Figure 35 Foam Parts 1 (upper foam) 1 (lower foam)
  • Page 177: Power And Status Light Pipes

    Parts and Materials Power and Status Light Pipes Power and Status Light Pipes Table 17 Power and Status Light Pipes Item Description Part Number Power switch coupler 5041-8383 Light pipe — power switch 5041-8382 Power switch button 5041-8381 Light pipe — status lamp 5041-8384 Figure 36 Power and Status Light Pipes...
  • Page 178: Leak Parts

    Parts and Materials Leak Parts Leak Parts Table 18 Leak Parts Item Description Part Number Leak sensor 5061-3356 Tube clip 5041-8387 Leak plane, pump 5041-8390 Leak funnel 5041-8388 Holder, leak funnel 5041-8389 Corrugated waste tube (reorder pack), 5m 5062-2463 Figure 37 Leak Parts...
  • Page 179: Pump-Head Assembly

    Parts and Materials Pump-Head Assembly Pump-Head Assembly Table 19 Pump-Head Assembly Item Description Part Number Complete assembly, included items marked with (*) G1311-60004 Sapphire plunger 5063-6586 Plunger housing (including springs) G1311-60002 Support ring 5001-3739 Seal (pack of 2) or 5063-6589 Seal (pack of 2), for normal phase applications 0905-1420 Capillary outlet valve to piston 2...
  • Page 180 Parts and Materials Pump-Head Assembly Figure 38 Pump-Head Assembly...
  • Page 181: Pump-Head Assembly With Seal Wash

    Parts and Materials Pump-Head Assembly with Seal Wash Pump-Head Assembly with Seal Wash Table 20 Pump-Head Assembly with Seal Wash Item Description Part Number Complete assembly, included items marked with (*) G1311-60005 Sapphire plunger 5063-6586 Plunger housing (including springs) G1311-60002 Support ring, seal wash 5062-2465 Secondary seal...
  • Page 182 Parts and Materials Pump-Head Assembly with Seal Wash Figure 39 Pump Head Assembly with Seal Wash Option...
  • Page 183: Outlet Ball Valve Assembly

    Parts and Materials Outlet Ball Valve Assembly Outlet Ball Valve Assembly Table 21 Outlet Ball Valve Assembly Item Description Part Number Outlet ball valve — complete assembly G1312-60012 Socket cap G1312-21208 Housing screw 01018-22410 Outlet valve cartridge No part number Sieve (pack of 10) 5063-6505 Gold seal, outlet...
  • Page 184: Purge-Valve Assembly

    Parts and Materials Purge-Valve Assembly Purge-Valve Assembly Table 22 Purge-Valve Assembly Item Description Part Number Purge valve — complete assembly G1311-60009 Valve body No part number PTFE frit (pack of 5) 01018-22707 Gold seal 5001-3707 Cap ( pack of 4, reorder number) 5062-2485 Figure 41 Purge Valve...
  • Page 185: Active Inlet Valve Assembly

    Parts and Materials Active Inlet Valve Assembly Active Inlet Valve Assembly Table 23 Active Inlet Valve Assembly Item Description Part Number Active inlet valve — complete assembly G1312-60010 Valve body No part number Valve cartridge 5062-8562 Figure 42 Active Inlet Valve Assembly Valve body Valve cartridge...
  • Page 186: Accessory Kit G1311-68705

    Parts and Materials Accessory Kit G1311-68705 Accessory Kit G1311-68705 Table 24 Tools and Accessories Description Part Number Wrench 14 mm 8710-1924 Insert tool seals 01018-23702 PTFE Frit (pack of 5) 01018-22707 Corrugated waste tube (1.2m) no PN Corrugated waste tube (reorder number), 5m 5062-2463 Velocity regulator 5062-2486...
  • Page 187: Seal Wash Option 01018-68722

    Parts and Materials Seal Wash Option 01018-68722 Seal Wash Option 01018-68722 Table 25 Seal Wash Option Description Part Number Support ring, seal wash (pack of 2) 5062-2465 Secondary seal (pre-installed in support ring) 0905-1175 Seal keeper (pack of 2) 5001-3743 Wash tube (1m) 0890-1764 5062-2486...
  • Page 188: Cable Overview

    Cable Overview WAR NI N G Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations. Table 26 Cables Overview Type Description Part Number Analog 3390/2/3 integrators 01040-60101...
  • Page 189 Agilent 1100 module to module, 1m lg 5181-1519 Agilent 1100 module to control module G1323-81600 External Agilent 1100 Series interface board to general G1103-61611 contacts purpose GPIB cable Agilent 1100 module to ChemStation, 1 m 10833A Agilent 1100 module to ChemStation, 2 m...
  • Page 190: Analog Cables

    Analog Cables Analog Cables One end of these cables provides a BNC connector to be connected to Agilent 1100 Series modules. The other end depends on the instrument to which connection is being made. Agilent 1100 to 3390/2/3 Integrators Connector...
  • Page 191 Parts and Materials Analog Cables Agilent 1100 to BNC Connector Connector 8120-1840 Agilent 1100 Signal Name Shield Shield Analog - Center Center Analog + Agilent 1100 to General Purpose Connector 01046-60105 3394/6 Agilent 1100 Signal Name Not connected Black Analog - Analog +...
  • Page 192: Remote Cables

    Remote Cables One end of these cables provides a Agilent Technologies APG (Analytical Products Group) remote connector to be connected to Agilent 1100 Series modules. The other end depends on the instrument to be connected to. Agilent 1100 to 3390 Integrators...
  • Page 193 Parts and Materials Remote Cables Agilent 1100 to 3392/3 Integrators Connector Active 01046-60206 3392/3 Agilent 1100 Signal Name (TTL) 1 - White Digital ground 2 - Brown Prepare run 3 - Gray Start 4 - Blue Shut down 5 - Pink Not connected 6 - Yellow Power on...
  • Page 194 Parts and Materials Remote Cables Agilent 1100 to 3396A Integrators Connector Active 03394-60600 3394 Agilent 1100 Signal Name (TTL) 1 - White Digital ground 2 - Brown Prepare run 3 - Gray Start 4 - Blue Shut down 5 - Pink Not connected 6 - Yellow Power on...
  • Page 195 Parts and Materials Remote Cables Agilent 1100 to 3396 Series III / 3395B Integrators Connector Active 03396-61010 33XX Agilent 1100 Signal Name (TTL) 1 - White Digital ground 2 - Brown Prepare run 3 - Gray Start 4 - Blue Shut down 5 - Pink Not connected...
  • Page 196 Parts and Materials Remote Cables Agilent 1100 to HP 1090 LC, HP 1040 DAD or Signal Distribution Module Connector Active 01046-60202 HP 1090 Agilent 1100 Signal Name (TTL) 1 - White Digital ground 2 - Brown Prepare run 3 - Gray Start 4 - Blue Shut down...
  • Page 197: Bcd Cables

    BCD Cables BCD Cables One end of these cables provides a 15-pin BCD connector to be connected to the Agilent 1100 Series modules. The other end depends on the instrument to be connected to. Agilent 1100 to 3392/3 Integrators Connector...
  • Page 198 Parts and Materials BCD Cables Agilent 1100 to 3396 Integrators Connector 03396-60560 3392/3 Agilent 1100 Signal Name BCD Digit BCD 5 BCD 7 BCD 6 BCD 4 BCD 0\ BCD 3 BCD 2 BCD 1 Digital ground + 5 V Agilent 1100 to General Purpose Connector 18594-60520...
  • Page 199: Auxiliary Cable

    Parts and Materials Auxiliary Cable Auxiliary Cable One end of this cable provides a modular plug to be connected to the Agilent 1100 Series vacuum degasser. The other end is for general purpose. Agilent 1100 Series Degasser to general purposes Connector G1322-81600...
  • Page 200: External Contact Cable

    External Contact Cable One end of this cable provides a 15-pin plug to be connected to Agilent 1100 Series module’s interface board. The other end is for general purpose. Agilent 1100 Series Interface Board to general purposes Connector G1103-61611 Color...
  • Page 201: Rs-232 Cable Kit

    RS-232 Cable Kit This kit contains a 9-pin female to 9-pin female Null Modem (printer) cable and one adapter. Use the cable and adapter to connect Agilent Technologies instruments with 9-pin male RS-232 connectors to most PCs or printers. Agilent 1100 module to PC...
  • Page 202: Lan Cables

    Parts and Materials LAN Cables LAN Cables Recommended Cables For point to point connection (not using a network hub) use a twisted pair cross over LAN cable (P/N 5183-4649, 10 feet long). For standard network connections using a hub use category 5 UTP cables, (P/N G1530-61480, 8 m long).
  • Page 203: Introduction To The Binary Pump

    Introduction to the Binary Pump An introduction to the pump, instrument overview, theory of operation, external communication and internal connectors...
  • Page 204: Introduction To The Binary Pump

    Introduction to the Binary Pump Introduction to the Binary Pump Introduction to the Binary Pump The binary pump comprises two identical pumps integrated into one housing. It provides gradient generation by high-pressure mixing. Degassing is not included but a vacuum degasser is available as a separate product for applications that require best flow stability especially at low flow rates or maximum detector sensitivity.
  • Page 205 Introduction to the Binary Pump Introduction to the Binary Pump Overview The binary pump is based on a two-channel, dual-piston in-series design which comprises all essential functions that a solvent delivery system has to fulfill. Metering of solvent and delivery to the high-pressure side are performed by two pump assemblies which can generate pressure up to 400 bar.
  • Page 206 Introduction to the Binary Pump Introduction to the Binary Pump How Does the Binary Pump Work? The liquid runs from the solvent reservoir through an active inlet valve. Each side of the binary pump comprises two substantially identical pump units. Both pump units comprise a ball-screw drive and a pump head with two sapphire pistons for reciprocating movement.
  • Page 207 Introduction to the Binary Pump Introduction to the Binary Pump range of 20 µl to 100 µl depending on the flow rate. The microprocessor controls all flow rates in a range of 1 µl/min to 5 ml/min. The inlet of the first pumping unit is connected to the active inlet valve which is processor-controlled opened or closed allowing solvent to be drawn into the first pump unit.
  • Page 208 Introduction to the Binary Pump Introduction to the Binary Pump During the drawing stroke of the first piston, the second piston delivers the drawn volume into the system. Table 27 Binary Pump Details Delay volume From mixing point to pump outlet, dependent on back pressure 180 –...
  • Page 209 Introduction to the Binary Pump Introduction to the Binary Pump How Does Variable Stroke Volume Work? Due to the compression of the pump-chamber volume each piston stroke of the pump will generate a small pressure pulsation, influencing the flow ripple of the pump.
  • Page 210: Electrical Connections

    • The CAN bus is a serial bus with high-speed data transfer. The two connectors for the CAN bus are used for internal Agilent 1100 Series module data transfer and synchronization. • One analog output provides a signal for integrators or data handling systems.
  • Page 211 Introduction to the Binary Pump Electrical Connections Figure 46 Electrical Connections to the Binary Pump Configuration switch GPIB Slot for interface board RS232 Remote Analog output CAN-Bus Power Plug Security Lever WAR NI N G To disconnect the binary pump from line, unplug the power cord. The power supply still uses some power, even if the power switch on the front panel is turned off.
  • Page 212: Instrument Layout

    Introduction to the Binary Pump Instrument Layout Instrument Layout The industrial design of the module incorporates several innovative features. It uses Agilent’s E-PAC concept for the packaging of electronics and mechanical assemblies. This concept is based upon the use of expanded polypropylene (EPP) layers foam plastic spacers in which the mechanical and electronic boards components of the module are placed.
  • Page 213: Early Maintenance Feedback (Emf)

    Introduction to the Binary Pump Early Maintenance Feedback (EMF) Early Maintenance Feedback (EMF) Maintenance requires the exchange of components in the flow path which are subject to mechanical wear or stress. Ideally, the frequency at which components are exchanged should be based on the intensity of usage of the instrument and the analytical conditions, and not on a predefined time interval.
  • Page 214 Introduction to the Binary Pump Early Maintenance Feedback (EMF) Using the EMF Counters The user-settable EMF limits for the EMF counters enable the early maintenance feedback to be adapted to specific user requirements. The wear of pump components is dependent on the analytical conditions, therefore, the definition of the maximum limits need to be determined based on the specific operating conditions of the instrument.
  • Page 215: The Electronics

    Introduction to the Binary Pump The Electronics The Electronics The electronics are comprised of four main components: • The high pressure pump main board (HPM), see page 216. • Power supply, see page 235. Optional: • Interface board (BCD/external contacts), see page 223. •...
  • Page 216 Introduction to the Binary Pump The Electronics High-Pressure Pump Main Board (HPM) The board controls all information and activities of all assemblies within the binary pump. The operator enters parameters, changes modes and controls the binary pump through interfaces (CAN, GPIB or RS-232C), connected to the user-interfaces.
  • Page 217 Information” on page 273. Interfaces The binary pump provides the following interfaces: • two CAN connectors as interface to other Agilent 1100 Series modules, • one GPIB connector as interface to the Agilent ChemStation, • one RS-232C as interface to a computer, •...
  • Page 218 Introduction to the Binary Pump The Electronics Figure 47 Block Diagram Pump Controller Board Motor drive B Encoder B Drive B Pump Motor B ϑ head B Motor control B Control Current control B Data/addr Active inlet Valve drive B valve B ASIC SSV A/ SSV B...
  • Page 219 Introduction to the Binary Pump The Electronics Figure 48 Block Diagram Pump Controller Board Multiplexer Filter & A/D converter Damper Pressure sensor amplifier Offset Analog output Integrator correction Pressure converter Amplifier & ambient measurement Leak sensor Drive Sensor Control signal Data/ address Diagnose signals (digital) Diagnose signals (analog)
  • Page 220 Introduction to the Binary Pump The Electronics Figure 49 Interconnection Diagram HPM Board +24V Inlet valve Electronic AIV B Z-panel drive B fuse +24V motor B Motor drive B +/-15V +36V encoder B Hardware switch +24V motor A Motor Processor drive A +/-15V encoder A...
  • Page 221: Firmware Description

    • a non-instrument specific section, called ‘resident system’, • an instrument specific section, called ‘main system’. Resident System This resident section of the firmware is identical for all Agilent 1100 series modules. Its properties are: • the complete communication capabilities (GPIB, CAN, LAN and RS-232C), •...
  • Page 222 Introduction to the Binary Pump Firmware Description Firmware Updates Firmware updates can be done using your user interface: • handheld control module with files from a PC-card or • Agilent ChemStation with files from floppy disk The file naming conventions are: xxxx-vvv.DLB, where xxxxis the product number, e.g.
  • Page 223: Optional Interface Boards

    Introduction to the Binary Pump Optional Interface Boards Optional Interface Boards The Agilent 1100 Series modules have one optional board slot that allows to add an interface board to the modules. Table 28 Optional Interface Boards Description Part Number BCD Board...
  • Page 224 Introduction to the Binary Pump Optional Interface Boards LAN Board The HP LAN board is actually an HP JetDirect card, which is a network interface card used in HP printers. N O T E One board is required per Agilent 1100 stack. If the Agilent 1100 stack has a DAD, then the DAD MUST be the module used for LAN board installation.
  • Page 225: Agilent 1100 Series Interfaces

    Introduction to the Binary Pump Agilent 1100 Series Interfaces Agilent 1100 Series Interfaces The Agilent 1100 Series modules provide the following interfaces: Table 30 Agilent 1100 Series Interfaces DA Detector Thermostatted MW Detector VW Detector Column Vacuum Interface Type Pumps...
  • Page 226 Introduction to the Binary Pump Agilent 1100 Series Interfaces WAR NI N G Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations, see “Cable Overview” on page 188.
  • Page 227 APG Remote Interface The APG Remote connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features as common shut down, prepare, and so on. Remote control allows easy connection between single instruments or systems to ensure coordinated analysis with simple coupling requirements.
  • Page 228 Introduction to the Binary Pump Agilent 1100 Series Interfaces Table 32 APG Remote Signal Distribution Signal Description DGND Digital ground PREPARE (L) Request to prepare for analysis (for example, detector lamp on). Receiver is any module performing preanalysis activities. START (L) Request to start run / timetable.
  • Page 229 Introduction to the Binary Pump Agilent 1100 Series Interfaces connector. The RS-232C is designed as DCE (data communication equipment) with a 9-pin male SUB-D type connector. The pins are defined as: Table 33 RS-232C Connection Table Direction Function Ground Figure 52...
  • Page 230: Setting The 8-Bit Configuration Switch

    Introduction to the Binary Pump Setting the 8-bit Configuration Switch Setting the 8-bit Configuration Switch The 8-bit configuration switch is located next to the GPIB connector. Switch settings provide configuration parameters for GPIB address, serial communication protocol and instrument specific initialization procedures. Figure 53 8-bit Configuration Switch Factory setting is shown...
  • Page 231 If you just want to change the GPIB address and need a detailed procedure, refer to the Installing Your ChemStation System handbook. Default GPIB address is set to the following addresses: Table 35 Default Addresses for Agilent 1100 Series Modules Module Address Binary Address...
  • Page 232 Introduction to the Binary Pump Setting the 8-bit Configuration Switch Use the following tables for selecting the setting which you want to use for RS-232C communication. The number 0 means that the switch is down and 1 means that the switch is up. Table 36 Communication Settings for RS-232C Communication Mode Select 1...
  • Page 233 Introduction to the Binary Pump Setting the 8-bit Configuration Switch Table 39 Parity Settings Switches Parity No Parity Odd Parity Even Parity Forced Cold-Start Settings Switches 1 and 2 do not force storage of this set of parameters in non-volatile memory.
  • Page 234 Introduction to the Binary Pump Setting the 8-bit Configuration Switch Switches 1 and 2 do not force storage of this set of parameters in non-volatile memory. Returning the switches 1 and 2 to other positions (other than being both up) will allow for normal operation. If you use the following switch settings and power the instrument up again, the instrument firmware stays in the resident part, that is, it is not operable as a binary pump.
  • Page 235: The Main Power Supply Assembly

    Introduction to the Binary Pump The Main Power Supply Assembly The Main Power Supply Assembly The main power supply comprises a closed assembly (no component-level repair possibility). The power supply provides all DC voltages used in the binary pump module. The line voltage can vary in a range from 100 –...
  • Page 236 Introduction to the Binary Pump The Main Power Supply Assembly No accessible hardware fuse is needed because the main power supply is safe against any short circuits or overload conditions on the output lines. When overload conditions occur, the power supply turns off all output voltages.
  • Page 237 Control Module Screens for the Binary Pump...
  • Page 238: Control Module Screens For The Binary Pump

    Control Module Screens for the Binary Pump This chapter is intended to introduce an operator to the screens available for operation of the Agilent 1100 binary pump with the Agilent 1100 control module. Please use the manual of the control module for further detailed reference. Major keys on the Agilent 1100 Control Module Return to previous screen and scroll through top layer views (Analysis, Settings)
  • Page 239: Screens Available From The Analysis Screen

    Control Module Screens for the Binary Pump Screens available from the Analysis screen Screens available from the Analysis screen The Analysis screen This is the wake-up screen, if the Agilent 1100 binary pump is the only configured Agilent 1100 module. It is used to enter the most common pump method parameters.
  • Page 240 Control Module Screens for the Binary Pump Screens available from the Analysis screen Setup View In the Setup view, modules can be added or removed to the view. Here, e.g. the autosampler, the thermostatted column compartment and the DA detector parameters are shown on the display as well. The number of parameters on the display is restricted as additional modules are added.
  • Page 241 Control Module Screens for the Binary Pump Screens available from the Analysis screen Pump ON/OFF From the Analysis screen use the F7 key to proceed to the turn on screen. Press F8 (On) once to turn on the pump. If more than one module is available, select the binary pump from the pop-up menu.
  • Page 242 Control Module Screens for the Binary Pump Screens available from the Analysis screen Settings Within the Settings you can change the pump parameters. You have access to a different set of parameters available through the F1-5 keys. F7 key resets the pump to default values.
  • Page 243 Control Module Screens for the Binary Pump Screens available from the Analysis screen Settings - Timetable With the F2 key (Timetable) you can list the timetable for the pump. Press F7 key (Insert) to add entries or F6 key (Delete) to remove entries. Use the F6 key (Done) to view the entered lines of the timetable.
  • Page 244 Control Module Screens for the Binary Pump Screens available from the Analysis screen Settings - Pressure With the F3 key (Pressure) you can change the settings for the pressure limits. Settings - Bottle With the F4 key (Bottle Fillings) you can adjust the settings for the bottle Fillings fillings to their current state.
  • Page 245 Control Module Screens for the Binary Pump Screens available from the Analysis screen Analog -Status Press F5 key (Views) and select Status. Status This is an example if an Agilent 1100 pump is configured standalone. Information on the actual flow rate , mobile phase composition, pressure and %-ripple, elapsed run time and the pressure plot are shown.
  • Page 246 Control Module Screens for the Binary Pump Screens available from the Analysis screen Press F6 key (Select). Here you can add additional online signals (maximum are 3). Additional signals could also be chromatograms or temperature signals from other modules. Use the Right/Left arrows to switch between Available and Selected Signals.
  • Page 247 Control Module Screens for the Binary Pump Screens available from the Analysis screen Use F2 key (PC-Card) to save a method on a PCMCIA card. Use the Right/Left arrows to switch between PC-Card and Instrument window. Use the UP/Down arrows to select the method. Use the F7/F8 keys (Copy) to enter available signals into the box for selected signals or vice versa.
  • Page 248: Screens Available From The System Screen

    Control Module Screens for the Binary Pump Screens available from the System screen Screens available from the System screen System screen Use the Esc key to receive Views on the F5 key. Choose System from the pull-down menu. This screen shows the last activities in the system. System - Control Use the F1 key (Control) to select the binary pump.
  • Page 249 Control Module Screens for the Binary Pump Screens available from the System screen System - On the System screen use the F2 key (Configure) to select the pump. Use Configuration the F1 key (Interfaces) to access the interface settings (if required). Press F4 (Bottle fillings) to adjust bottle fillings to the currrent state.
  • Page 250: Screens Available From The Records Screen

    Control Module Screens for the Binary Pump Screens available from the Records screen Screens available from the Records screen Records screen Use the Esc key to receive Views on the F5 key. Choose System from the pull-down menu. Use the F4 key (Records) to select the pump. Errors are reported either into the System Log (F2) or Error Log (F3).
  • Page 251 Control Module Screens for the Binary Pump Screens available from the Records screen Info Log Use the m-key to receive a pop-up menu, Select Info Log. A list of the last events are listed. For troubleshooting reasons they can be printed or saved to a file on the PCMCIA card (using the m-key for the context sensitive menu).
  • Page 252 Control Module Screens for the Binary Pump Screens available from the Records screen If a set limit has been exceeded, a message box will pop up. This will not stop a sequence or run (information only to plan maintenance activities). If you press Reset, the limits will be removed.
  • Page 253 Control Module Screens for the Binary Pump Screens available from the Records screen Use the Esc key to receive Views on the F5 key. Choose System from the pull-down menu. Use the F3 key (Records) to select the Generic module. In this screen the resident firmware revision is shown.
  • Page 254 Control Module Screens for the Binary Pump Screens available from the Records screen Changing the serial In case the serial number of the module has to be added, use the m-key to number open the menu Enter Serial#. The serial number becomes active after restart of the module.
  • Page 255 Control Module Screens for the Binary Pump Screens available from the Records screen Use the F7 key (Add) to add new maintenance activities. If an activity is not listed, you can type the activity into the line “Add” using the control modules key pad.
  • Page 256: Diagnostics And Tests

    Control Module Screens for the Binary Pump Diagnostics and Tests Diagnostics and Tests Tests screen Use the Esc key to receive Views on the F5 key. Choose System from the pull-down menu. Use the F3 key (Tests) to select the pump. Two tests are available to test the Agilent 1100 pumps.
  • Page 257 Control Module Screens for the Binary Pump Diagnostics and Tests Pressure Test Use the F2 key (Pressure Test) to perform a pressure test of the system. Several steps like purging the system, setting up Isopropanol as solvent on channel A or A2 (if solvent selection valve is installed) and blocking the column outlet with a blank nut have to be performed before operating the pressure test.
  • Page 258 Control Module Screens for the Binary Pump Diagnostics and Tests...
  • Page 259 Specifications Performance specifications of the binary pump...
  • Page 260: Specifications

    Specifications Performance Specifications Table 43 Performance Specification Agilent 1100 Series Binary Pump Type Specification Hydraulic system Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, floating piston design and active inlet valve Setable flow range Setpoints 0.001 – 5 ml/min, in 0.001 ml/min increments Flow range 0.1 –...
  • Page 261 Specifications Performance Specifications Table 43 Performance Specification Agilent 1100 Series Binary Pump Analog output For pressure monitoring, 2 mV/bar, one output Communications Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional Safety and maintenance...
  • Page 262 Specifications Performance Specifications...
  • Page 263 Index overview, 188 electronic fuses, 210, 217 accessory kit, 18, 186 electronics, HPM board, 215 power, 17 active inlet valve, 108, 109, 185 remote, 18, 188, 192 electrostatic discharge (ESD), 104 active inlet valve drive, 216 signal, 18 adapter, 28, 109 CAN bus, 210 on control module, 251 Agilent on internet, 276...
  • Page 264 Index RS-232C, 227 active inlet valve, 108, 109 frit, 114 damper, 143 fuse, 210, 217 interface board, 131 fan, 145 interface cable, 22 high pressure pump main board interfaces, 217 gas solubility, 34 internet, 276 (HPM board), 136 introduction to the pump, 204 interface board, 131 GPIB internal parts, 104, 132...
  • Page 265 Index RS-232 non-operating temperature, 16 power supply, 151, 235 not-ready condition, 49 power supply indicator, 49 cable kit to PC, 201 power supply specification, 236 RS-232C communication setting, 231 power switch, 21 RS-232C connection table, 229 power-input socket, 14 RS-232C connector, 210 onboard battery, 217 RS-232C interface, 227 operating altitude, 16...
  • Page 266 Index solvent filters velocity regulator, 18, 40 checking, 37 voltage range, 16, 210, 236 cleaning, 38 voltage selector, 210 solvent information, 36, 275 solvent inlet filter, 34 solvent selection valve, 24, 204 wait timeout, 84 solvent selection valve drive, 216 wall socket, 14 solvents, 275 warranty...
  • Page 267: Warranty Statement

    Warranty Statement Warranty Statement All Chemical Analysis Products Agilent Technologies warrants its chemical analysis products against defects in materials and workmanship. For details of the warranty period in your country, call Agilent. During the warranty period, Agilent will, at its option, repair or replace products which prove to be defective.
  • Page 268 Warranty Statement 4 operation outside of the environmental and electrical specifications for the product, 5 improper site preparation and maintenance, or 6 customer induced contamination or leaks. THE WARRANTY SET FORTH IS EXCLUSIVE AND NO OTHER WARRANTY, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED. AGILENT SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
  • Page 269 Warranty Statement Responsibilities of Agilent Technologies Agilent Technologies will provide warranty services as described in Table 44. Table 44 Warranty Services Services During Warranty Warranty Period Type Agilent 1100 Series of Modules 1 Year Onsite GC, LC, UV-Visible, and LAS supplies and...
  • Page 270: Safety Information

    Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. General...
  • Page 271 Safety Information rendering first aid and resuscitation, is present. Do not replace components with power cable connected. Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
  • Page 272 Safety Information CA UT IO N A caution alerts you to situations that could cause a possible loss of data. Do not proceed beyond a caution until you have fully understood and met the indicated conditions.
  • Page 273: Lithium Batteries Information

    Lithium Batteries Information Lithium Batteries Information WAR NI N G Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended by the equipment manufacturer. Lithium batteries may not be disposed-off into the domestic waste. Transportation of discharged Lithium batteries through carriers regulated by IATA/ICAO, ADR, RID, IMDG is not allowed.
  • Page 274: Radio Interference

    Radio Interference Radio Interference Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations. Test and Measurement If test and measurement equipment is operated with equipment unscreened cables and/or used for measurements on open set-ups, the user has to assure that under operating conditions the radio interference limits are still met within the premises.
  • Page 275: Solvent Information

    Solvent Information Solvent Information Observe the following recommendations on the use of solvents. Flow Cell Avoid the use of alkaline solutions (pH > 11) which can attack quartz and thus impair the optical properties of the flow cell. Solvents Always filter solvents through 0.4 µm filters, small particles can permanently block the capillaries and valves.
  • Page 276: Agilent Technologies On Internet

    Agilent Technologies on Internet For the latest information on products and services visit our worldwide web site on the Internet at: http://www.agilent.com/go/chem Select “Products” - “Chemical Analysis” It will also provide the latest firmware of the Agilent 1100 series modules for download.
  • Page 278 In This Book This manual contains technical reference information about the Agilent 1100 Series binary pump. The manual describes the following: • installation, • optimizing performance, • diagnostics and troubleshooting, • repairing, • parts and materials, • introduction to the pump, theory of operation, •...

Table of Contents