Match The Flow Cell To The Column - Agilent Technologies 1200 Infinity Series User Manual

Hide thumbs Also See for 1200 Infinity Series:
Table of Contents

Advertisement

Match the Flow Cell to the Column

The tables below recommend the flow cell that matches the column used.
If more than one selection is appropriate, use the larger flow cell to get
the best detection limit. Use the smaller flow cell for best peak resolution.
Standard HPLC Applications
Figure 34
Ultra fast separation with RRLC systems
Figure 35
• (+) For ultra fast analysis with step gradients the micro flow cell (2 µL,
• (++) In high resolution analysis time is not the highest priority. Higher
• If longer columns (> 50 mm) for higher resolution are used, then the
Agilent 1200 Infinity Series VWD - User Manual
Choosing a Flow Cell (Standard HPLC Applications)
Choosing a Flow Cell for G1314E (for ultra fast separation with RRLC systems)
3 mm) gives the best performance
delay volumes are accepted. Therefore we recommend to use the
damper plus mixer for a highest signal to noise.
next larger flow cell is the preferred choice for higher sensitivity.
How to optimize the detector
Match the Flow Cell to the Column