Replacing Capillaries On A Semi-Micro And High Pressure Flow Cell - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Replacing Capillaries on a Semi-Micro and High Pressure Flow
Cell
When required
Tools required
Parts required
Preparations for this procedure:
Turn off the flow.
Remove the front cover.
Remove the flow cell, see
page 105.
1100 Series DAD and MWD Reference Manual
If the capillary is blocked
Two 1/4 inch wrenches for capillary connections
Wrench 4 mm for capillary connections
Screwdriver Pozi 1 PT3
For parts see
"Exchanging a Flow Cell"
"Semi-Micro Flow Cell"
1 Identify the inlet and outlet capillaries.
Inlet
on
capillary
Repairing the Detector
on page 188
Outlet capillary
4
119