Stay-Resident Settings

Firmware update procedures may require this mode in case of firmware
loading errors.
Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning switches 1 and 2 to other positions (other than being both
up) will allow for normal operation.
If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident part, that is, it is not operable as
a detector. It only uses basic functions of the operating system for example, for
communication.
Table 47
Mode Select
TEST/BOOT
To return to normal operation, set switches back to your GPIB or RS-232C
configuration settings.
1100 Series DAD and MWD Reference Manual
Stay Resident Settings
1
2
3
1
1
0
Introduction to the Detector
4
5
6
0
1
0
6
7
8
0
0
265