Flow Cell Flushing

When
If flow cell is contaminated
Tools required
Glass syringe, adapter
Parts required
#
1
Dangerous solvents
WA R N I N G
The strong solvents used in this procedure are toxic and flammable and proper
precautions are necessary.
Aqueous solvents in the flow cell can build up algae. Therefore do not leave aqueous
N O T E
solvents in the flow cell for long periods. Add a small percentage of organic solvents (e.g.
Acetonitrile or Methanol ~ 5%).
The strong solvent should dissolve any potential contaminants in the flow cell. For example
N O T E
water for aqueous mobile phase buffers, chloroform or tetrahydrofuran for not water
soluble contaminants.
In case the cell is contaminated, follow the procedure below.
1 Flush with the strong solvent.
2 Leave this solution in the cell for about one hour.
3 Flush with mobile phase.
Do not exceed the flow cell pressure limit of 5 bar (0.5 MPa).
N O T E
Agilent 1260 Infinity RID User Manual
Description
Strong solvent, tubings to waste
Wear protective gloves and goggles.
Don't expose yourself to the vapors.
9
Maintenance

Flow Cell Flushing

135