7
Error Information
Detector Error Messages
Flow Cell Removed
The detector has an automatic cell recognition system. When the flow cell is
removed, the lamp is turned off and a NOT READY condition exists. If the flow
cell is removed during an analysis, a SHUT DOWN is generated.
Probable cause
1
No Peaks
If no peaks are shown in the chromatogram, the user-interface shows the
module still in "Ready". There is no feedback mechanism that checks whether
the lamp is ON.
Probable cause
1
2
3
4
5
142
Flow cell has been removed during analysis.
Lamp is off.
Defective FLL board / Trigger pack.
Defective Xenon flash lamp.
Wrong position of monochromator.
Defective FLF board.
Suggested actions
Insert flow cell and turn on the lamp.
Suggested actions
Perform a "Lamp Intensity Test" (see
Intensity Test"
on page 149). If no profile
available (very low counts).
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Perform a "Wavelength Accuracy Test" (see
"Wavelength Accuracy Test"
check the wavelength calibration.
Please contact your Agilent service
representative.
Agilent 1260 FLD User Manual
"Lamp
on page 167) to