Replacing Leak Handling System Parts

When required
Tools required
Parts required
1 Remove the front cover.
2 Pull the leak funnel out of the leak funnel holder.
3 Pull out the leak funnel with the tubing.
4 Insert the leak funnel with the tubing in its position.
5 Insert the leak funnel into the leak funnel holder.
6 Replace the front cover.
Leak funnel holder
Leak funnel
Leak tubing
Figure 32
1100 Series DAD and MWD Reference Manual
If the parts are corroded or broken
None
Leak funnel 5061-3356
Leak funnel holder 5041-8389
Leak tubing (120 mm) 0890-1711
Replacing Leak Handling System Parts
Repairing the Detector