Table 1
Flow Cell Data
STD
Maximum pressure
40 (4)
Path length
10 (conical)
Volume
14
Inlet i.d.
0.25
Inlet length
750
Outlet i.d.
0.30
Outlet length
120
Total volume
60.77
Materials in contact
SST, quartz, PTFE,
with solvent
PEEK
Agilent InfinityLab LC Series VWD User Manual
Flow Cell with RFID tag
Semi-micro
40 (4)
6 (conical)
5
0.17
250
0.17
120
14.49
SST, quartz, PTFE
RFID tag
Micro
High Pressure
120 (12)
400 (40)
3 (conical)
10 (conical)
2
14
0.12
0.25
310
750
0.17
0.17
120
120
14.00
60.77
SST, quartz, PTFE
SST, quartz, Kapton
bar (MPa)
mm
µL
mm
mm
mm
mm
µL
14