Agilent Technologies 1100 Series Reference Manual page 297

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Table 50
Performance Specifications Agilent 1100 Series DAD and MWD, continued
Type
Control and data
evaluation
Analog outputs
Communications
Safety and maintenance Extensive diagnostics, error detection
GLP features
Housing
ASTM: "Standard Practice for Variable Wavelength Photometric Detectors Used in Liquid
N O T E
Chromatography".
Reference conditions: cell path length 10 mm, response time 2 s, flow 1 ml/min LC-grade
Methanol, slit width 4 nm.
Linearity measured with caffeine at 265 nm.
For environmental conditions refer to
1100 Series DAD and MWD Reference Manual
Specification
Agilent ChemStation for LC
Recorder/integrator: 100 mV or 1 V,
output range 0.001 – 2 AU, two
outputs
Controller-area network (CAN), GPIB,
RS-232C, APG Remote: ready, start,
stop and shut-down signals, LAN
optional
and display (through control module
and ChemStation), leak detection, safe
leak handling, leak output signal for
shutdown of pumping system. Low
voltages in major maintenance areas.
Early maintenance feedback (EMF) for
continuous tracking of instrument
usage in terms of lamp burn time with
user-settable limits and feedback
messages. Electronic records of
maintenance and errors. Verification of
wavelength accuracy with built-in
holmium oxide filter.
All materials recyclable.
"Environment"
Specifications
Comments
on page 15.
8