7 Remove the detector main board. Place the board on the ESD kit.
8 In most cases the RFI spring plate remains on the interface connectors of
9 On the new board check the switch setting of address switch S1, see
An incorrect switch setting (for example, TEST/BOOT) may cause the module to turn in a
N O T E
basic mode (yellow or red flashing status light). In such a case turn off the module, reset
the address switches, and turn on the module again.
10 Install the new detector main board and reconnect the connectors. Assure
11 Refit the screws at the REMOTE and GPIB connectors and the nuts of the
12 Reinstall the top foam section, top cover and front cover, see
13 Replace detector into the stack and reconnect the cables.
If a new detector main board is installed, update the serial number information of the
N O T E
detector in the user interface, see procedure below.
14 Check the firmware revision of the module. If the firmware revision is older
The G1315B DAD and the G1365B MWD require firmware revision 4.x or above (main and
N O T E
resident).
1100 Series DAD and MWD Reference Manual
the board. Carefully remove the spring plate and place it back into its
position in the instrument before installing a new board.
"Setting the 8-bit Configuration Switch"
that the board is fitted correctly into the board recess in the rear panel.
ANALOG connectors.
Foam and the Top Cover"
than the current firmware revision of the module, update the firmware
using the standard firmware update procedure, see
Detector's Firmware"
on page 261.
on page 175.
on page 180.
Repairing the Detector
"Replacing the
"Replacing the
4
145