4
Repairing the Detector
2 Carefully insert the optical unit into the instrument holding
the one hand at back of the heat sink.
4 Replace the two rubber shock absorbers and reconnect
the ground connector.
If a new optical is installed, perform a
N O T E
recalibration parameters into the detector main board's memory after turning on the
detector.
172
3 Reconnect the SCI cable (J23) and the holmium oxide filter
motor (J19) to the detector main board.
Next steps:
•
Reinstall the fan assembly (J16). Ensure the correct
direction of air flow.
•
Reinstall the flow cell cover, see
Exchanging the Holmium Oxide Filter"
•
Reinstall the flow cell, see
page 105.
•
Reinstall the lamps, see
page 102.
•
Reinstall the front cover, top cover and top foam section,
see
"Replacing the Foam and the Top Cover"
page 175.
•
Replace detector into the stack.
•
Reconnect the power cable and turn on the detector.
"Wavelength Recalibration"
1100 Series DAD and MWD Reference Manual
"Cleaning or
on page 127.
"Exchanging a Flow Cell"
"Exchanging a Lamp"
on
on
on page 84 to add the
on