Agilent Technologies 1100 Series Reference Manual page 172

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector
2 Carefully insert the optical unit into the instrument holding
the one hand at back of the heat sink.
4 Replace the two rubber shock absorbers and reconnect
the ground connector.
If a new optical is installed, perform a
N O T E
recalibration parameters into the detector main board's memory after turning on the
detector.
172
3 Reconnect the SCI cable (J23) and the holmium oxide filter
motor (J19) to the detector main board.
Next steps:
Reinstall the fan assembly (J16). Ensure the correct
direction of air flow.
Reinstall the flow cell cover, see
Exchanging the Holmium Oxide Filter"
Reinstall the flow cell, see
page 105.
Reinstall the lamps, see
page 102.
Reinstall the front cover, top cover and top foam section,
see
"Replacing the Foam and the Top Cover"
page 175.
Replace detector into the stack.
Reconnect the power cable and turn on the detector.
"Wavelength Recalibration"
1100 Series DAD and MWD Reference Manual
"Cleaning or
on page 127.
"Exchanging a Flow Cell"
"Exchanging a Lamp"
on
on
on page 84 to add the
on