Agilent Technologies 1100 Series Reference Manual page 166

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector
9 Insert the power supply into its location and fix it with the
screws at the rear panel.
11 Reinstall bottom foam part. Slide it in underneath the leak
drain.
166
10 Press down and clip in the power switch light pipe into the
power supply.
Power
switch light
pipe
Next steps:
Reinstall the processor board, see
Detector Main Board"
Reinstall the fan assembly, see
(G1315A/G1365A)"
on page 148 or
(G1315B/G1365B)"
on page 150.
Reinstall the optical unit, see
(G1315A/G1365A)"
on page 171 or
Optical Unit (G1315B/G1365B)"
Reinstall the front cover, top cover and top foam section,
see
"Replacing the Foam and the Top Cover"
page 175.
Replace detector into the stack.
Reconnect the power cable and turn on the detector.
1100 Series DAD and MWD Reference Manual
"Exchanging the
on page 143.
"Exchanging the Fan
"Replacing Fan Parts
"Installing the Optical Unit
"Removing the
on page 156.