4
Repairing the Detector
Replacing Fan Parts (G1315B/G1365B)
With the introduction of the G1315B/G1365B DAD/MWD a heater and a temperature
N O T E
sensor was added to the standard fan assembly. Also the air flow direction is reversed
(now: rear to front). Use this procedure for the the G1315B/G1365B only. For
G1315A/G1365A detectors refer to
When required
Tools required
Parts required
Preparations for this procedure:
•
Turn off the lamp.
•
Switch off the detector, and disconnect the cables and
capillaries.
•
Remove detector from stack and place it on the working
bench.
•
Remove front cover, top cover and top foam section, see
"Removing the Top Cover and Foam"
150
"Exchanging the Fan (G1315A/G1365A)"
If the fan is defective or noisy, if the temperature sensor or the heater fails
Screwdriver POZI 1 PT3, a pair of pliers
Fan assembly 3160-1016,
Temperature sensor assembly G1315-60000,
Heater assembly G1315-60003
1 Disconnect the connectors from the main board; fan (J16),
on page 140.
temperature sensor (J18) and heater (J17) and pull the fan
assembly out of its location.
1100 Series DAD and MWD Reference Manual
on page 148.