Shutdown

An external instrument has generated a shut-down signal (through CAN or
REMOTE lines).
The detector continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
• Leak detected in an external instrument with a remote connection to the
• Shut-down in an external instrument with a remote connection to the
• The degasser failed to generate sufficient vacuum for solvent degassing.
1100 Series DAD and MWD Reference Manual
Probable Causes
system.
system.
Suggested Actions
Fix the leak in the external instrument before restarting the pump.
Check external instruments for a shut-down condition.
Check the degasser for an error condition. Refer to the Reference Manual
for the Aligent 1100 Series vacuum degasser.
Troubleshooting and Test Functions