Repairing the Detector
6
Disconnect the lamps from their connectors.
8
While removing the top foam section completely, route the
lamp cables through the hole in the foam.
142
7
Remove the two lamp connectors towards the back by
depressing their connector tabs with a screwdriver.
Note:
Do not connect a power plug to the detector module
after removing the top covers.
A safety light switch on the main board will turn off
fan (immediately) and electronics (after 30 s) to
avoid the operation with removed covers. An error
will be generated (status lamp will light red) and the
logbook will show an error message.