12 Before installing the filter motor, press the filter lever in
from the other side.
14 Slide in the cell support assembly until it is stopped by the
holmium filter. Leave about 0.2-mm clearance for
movement of the filter lever between both the lens
assembly and the filter lever, and fix the setscrew.
1100 Series DAD and MWD Reference Manual
13 While counter-holding the filter lever, insert the filter
motor onto the filter shaft until it clicks into the final
position.
Next steps:
Check that the filter lever is moveable.
Reinstall the optical unit, see
(G1315A/G1365A)"
on page 171 or
Optical Unit (G1315B/G1365B)"
Reinstall the flow cell, see
page 105.
Reinstall the front cover, top cover and top foam section,
see
"Exchanging the Detector Main Board"
Replace detector into the stack.
Reconnect the cables and turn on the detector.
Repairing the Detector
"Installing the Optical Unit
"Removing the
on page 156.
"Exchanging a Flow Cell"
on page 143.