5 Fix the new heat exchanger to the clamp unit and the heat
exchanger to the cell body.
Next steps:
•
Reconnect the capillaries, see
•
Perform a leak test.
•
Insert the flow cell.
•
Replace the front cover.
•
Perform a wavelength calibration, see
Test"
on page 86 to check the correct positioning of the flow cell.
1100 Series DAD and MWD Reference Manual
"Exchanging a Flow Cell"
"Wavelength Recalibration"
6 Fix the inlet capillary to the flow cell body handtight first.
Then do a 1/4 turn with a 4-mm wrench.
on page 105.
on page 84 or holmium test, see
Repairing the Detector
Inlet capillary
"Holmium Oxide
4
121