How to optmize the Detector
Table 4
Optimization Overview, continued
Parameter
5 Setting the slit width
Use 4 nm slit for normal applications.
Use narrow slit (e.g 1 nm) if your analytes have narrow absorbance
bands and for high concentrations.
Use a wide slit (e.g. 16 nm) to detect very low concentrations.
Optimizing spectral acquisition (DAD only)
Select spectra acquisition mode according to your needs (see
for storing spectra"
on page 42).
Set the spectral wavelength range (for colorless samples 190...400 nm
is sufficient).
Set step to 4 nm for normal use; set small step (and slit width) if high
resolution of spectra with fine structure is wanted.
Typical column
Typical peak
length
width
T<= 5 cm
0.025 min
10 cm
0.05 min
20 cm
0.1 min
>= 40 cm
0.2 min
Typical flow
rate
Internal column diameter
Figure 7
32
Recommended flow cell
500 nl
flow cell
Semi-mirco
flow cell
0.2 ... 0.4
0.01 ... 0.2
ml/min
ml/min
0.5 ... 1 mm
2.1 mm
Choosing a Flow Cell
Impact
• spectral resolution, sensitivity and
linearity.
"Options
Standard
flow cell
1 ... 2 ml/min
0.4 ... 0.4
ml/min
3.0 mm
4.6 mm
1100 Series DAD and MWD Reference Manual
High pressure
flow cell for
pressures
above 100 bar
0.01 ... 5
ml/min