5
Replace the flow cell cover and fix the six screws.
1100 Series DAD and MWD Reference Manual
Next steps:
•
Perform a holmium oxide test, see
Test"
on page 86 to check the proper function of the
holmium oxide filter.
•
Insert the flow cell, see
page 105.
•
Replace the front cover.
•
Turn on the flow.
Repairing the Detector
"Holmium Oxide
"Exchanging a Flow Cell"
4
on
129