9
Maintenance and Repair
Exchanging a Flow Cell
Exchanging a Flow Cell
When
If application needs a different type of flow cell or the flow cell needs repair.
Tools required
Description
Wrench, 1/4 inch
for capillary connections
Parts required
#
1
For flow cell details see
•
•
•
•
Preparations
Turn the lamp OFF.
1
Press the release buttons and remove the front cover to
have access to the flow cell area.
138
Description
Flow cell
"Standard Flow Cell 10 mm / 14 µL"
"Micro Flow Cell 3 mm / 2 µL"
"Semi-micro Flow Cell 6 mm / 5 µL"
"High Pressure Flow Cell 10 mm / 14 µL"
on page 152
on page 154
on page 156
on page 158
2
Disconnect the inlet and outlet capillaries.