Pressure Plate Adjustment; Connecting The Power And The Communications Cable - Agilent Technologies SPS 4 User Manual

Autosampler
Table of Contents

Advertisement

3
Installation
• Keep delivery and suction lines as short as possible.
N O T E
• Run at a slow speed when pumping viscous fluids.
• Use the largest bore tube running at a slow speed for the longest tube life.
• It is recommended to release the occlusion arm when the pump is not in use to extend

Pressure Plate Adjustment

The position of the occlusion spring on each channel of the wash pump is
set in the factory to optimize the tension on the pressure plate when
pump tubing is fitted to the wash pump. It is possible to adjust the
tension on the pressure plate on a specific channel (only if the wash
pump can be turned on via the instrument control software) using the
following procedure:
1 Fit a new pump tube to the specific wash pump channel as described
2 Connect the inlet line to the wash reservoir and place the outlet line
3 Loosen the occlusion screw with a 5.5 mm (7/32") wrench until there is
4 Turn the wash pump on using the instrument control software.
5 While the wash pump is running, tighten the occlusion screw until rinse
6 Tighten the screw a further half turn.

Connecting the Power and the Communications Cable

Before connecting any part of the autosampler to the mains power supply,
check you have the correct power supply. The power requirements are
given in
48
the life of the pump tubing.
on
Page
47.
into a beaker.
no pressure on the pump tubing.
solution starts to flow.
"Power"
on Page 22.
Agilent SPS 4 Autosampler User's Guide