Troubleshooting and Error Handling

Power Failure on AI

Possible Causes
The high performance autosampler was switched
off
The 24V power was disconnected from the high
performance autosampler or from the
Automation Interface
The 24V power cord is damaged
The internal 24 V power circuit has failed

High Performance autosampler Thermostat Enabled Error

Possible Causes
The high performance autosampler thermostat
was switched on
94
Actions
Switch ON the high performance autosampler
Reconnect the 24V power cord
Replace the 24V power cord
Verify functionality using Automation Interface
test in diagnosis
Actions
The high performance autosampler thermostat
should not be used while the Automation
Interface is installed