REPLACING THE SOFTWARE

5 REPLACING THE SOFTWARE

5.1 Keyboard processor PCB software

NOTE

5.2 Generator processor PCB software

Technical Manual
Remove the lower shelf cover as described in section 1.5 "Lower shelf cover" on page H-4.
Remove the software chip from its socket with the special tool as illustrated on Fig. 6 below.
Figure 6
Carefully place the new software chip into the socket.
The orientation of the software chip is critical. Never try to force the chip into the
socket.
Attach the lower shelf cover to its position.
Remove the tube head inner cover and the rotating arm cover as described in sections 1.3
"Tube head covers" on page H-3 and 1.4 "Rotating arm cover" on page H-3.
Detach all the cables from the generator processor PCB.
Chapter H - PARTS REPLACEMENT & REPAIR