Patient Positioning Mechanism Adjustment - Planmeca Proline XC Technical Manual

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

PATIENT POSITIONING MECHANISM

3.2 Patient positioning mechanism adjustment

Technical Manual
If the patient positioning mechanism is too near or too far from the column it will have to be
centered.
Remove the cover from the lower shelf. Attach the ball phantom tool.
Select the Test mode as follows. Touch the i-field on the Main display. The list of User set-
tings appears.
Figure 40
From the list of User settings select Behavioural preferences . The display shown below
appears. Touch the Test mode field to enable the test mode (ON).
Figure 41
Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline XC X-ray unit D-21