CEPHALOSTAT ADJUSTMENT
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
In case the triangle is not evenly shaped, it can be slightly edited. Enlarge the image if
needed. The sensor joint is marked with the red lines when the keyboard's L-key is pressed.
Move the lower part of the triangle with the keyboard's arrow keys as follows: the right edge of
the triangle (vertical) can be adjusted with the right/left arrow keys and the left edge with the
up/down keys.
Enlarge the image
Red lines
Figure 156
While editing the shape of the triangle the sensor is simultaneously recalibrated.
If the evenly shaped triangle can not be reached, the second primary collimator position must
be adjusted in vertical direction. Refer to the section 7.8 "Second primary collimator adjust-
ment" on page E-78 for information on how to align it.
Checking the calibration
You can test whether the calibration succeeded as follows. Do not remove the calibration
block from the sensor head.
Take a test exposure by selecting Test Ceph Cal File from the Calibrate menu.
Figure 157
Planmeca Proline XC X-ray unit E-83
Technical Manual