CEPHALOSTAT ADJUSTMENT
4
PLANMECA PM 2002 Proline Cephalostat 18x24
Technical Manual
The radiation beam must appear within the borders of the rectangle marked on the alignment
tool. If it does not refer to the section 2.3 "Primary collimator adjustment, Planmeca Proline
XC Pan/Ceph X-ray unit" on page D-12. The adjustment is performed in the same way as
when adjusting the standard panoramic collimator.
Beam
5
No. 652232
Figure 89
Turn the beam alignment tool so that it is horizontal.
Cassette
holder
Beam
alignment
tool
Figure 90
Select the collimator number 6.
Figure 91
Chapter D - ADJUSTMENT & CALIBRATION
6
Positioning light 25mm
Positioning light 0mm
Soft tissue filter 0mm
Planmeca Proline XC X-ray unit D-55
Fifth primary collimator