Planmeca Proline XC Technical Manual page 156

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
From the list of Service settings select Patient positioning adjustment.
Figure 94
The display shown below appears. Adjust the position of the patient positioning mechanism
with the arrow fields until the line on the ball phantom lines up with the line on the alignment
ruler. The position of the mechanism is shown above the arrows. Accept the new value by
touching the OK field.
Figure 95
Remove the alignment pin and exit the service settings mode.
Take another ball phantom picture to check the alignment again. Readjust if necessary.
E-46 Planmeca Proline XC X-ray unit
PATIENT POSITIONING MECHANISM
Technical Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents