Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
If the distances from the middle ball to the center of the tenth ball on both the right and left are
more than 122mm then the patient positioning mechanism is too far backward and must be
moved forwards (Fig. 82).
Right-hand side
Figure 82
If the distances from the middle ball to the center of the tenth ball on both the right and left are
less than 118mm then the patient positioning mechanism is too far forward and must be
moved backwards (Fig. 83).
Right-hand side
Figure 83
E-40 Planmeca Proline XC X-ray unit
PATIENT POSITIONING MECHANISM
More than 122 mm More than 122 mm
Less than 118 mm Less than 118 mm