Planmeca Proline XC Technical Manual page 135

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

PANORAMIC X-RAY BEAM ADJUSTMENT
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
The system suggests default kV and mA values. The default kV value for the beam check is
60 and the default mA value is 4. Touch the Ready field.
Figure 46
NOTE
In case the contrast on the exposure is not satisfactory, change the values by touching
the kV/mA field and take a test exposure.
The following display appears on the computer screen.
Figure 47
Press the X-ray unit's exposure switch.
After the exposure is taken the following window appears.
Figure 48
You can type any kind of additional information (e.g. name of the clinic) on this window. Con-
firm by clicking the OK button.
Planmeca Proline XC X-ray unit E-25
Technical Manual