CEPHALOSTAT ADJUSTMENT
7.3
Checking the cephalometric beam alignment
NOTE
Technical Manual
Select the Cephalostat program as follows. Touch the Prog. field on the main display.
Figure 122
The Select program type appears. Select Ceph. The rotating unit will move automatically to
the correct position for taking cephalometric exposures.
Figure 123
Select kilovolt and milliampere values high enough to enable the radiation beam to be seen in
a darkened room for example 80kV and 12mA.
In case the second primary collimator is attached to the cephalostat, remove it before
checking the ceph beam alignment.
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline XC X-ray unit E-65