PATIENT POSITIONING LIGHTS
5.2 Positioning light adjustment
Technical Manual
Bending the mirror bracket backwards or forwards will position the light beam.
Figure 58
Rotating the mirror bracket will adjust the light beam in vertical direction.
Figure 59
Chapter D - ADJUSTMENT & CALIBRATION
Mirror bracket
Planmeca Proline XC X-ray unit D-33