PROBLEMS IN RADIOGRAPHS
10.3 Exposure on - no picture
Film unexposed
Film completely black
10.4 Very light radiographs
Very light radiographs
10.5 Stripes on the film
Horizontal stripes
Vertical stripes
Technical Manual
SYMPTOM
SYMPTOM
SYMPTOM
POSSIBLE REASON
Primary collimator does not let
in radiation to film.
In the developer the fixative
and developer changed places.
Light leakages to dark room.
Film otherwise exposed.
POSSIBLE REASON
Film cassette upside down in
its holder.
Primary collimator positioned
so that the radiation beam goes
trough the secondary collima-
tor.
Secondary collimator moved.
Wrong
film/intensifying
combination.
POSSIBLE REASON
From developing machine.
Primary collimator tilted so that
part of the film meets second-
ary collimator.
Patient has jewels, large fillings
in teeth, etc.
From
rolls
of
developing
machine.
Cassette mechanism gets tuck.
Generator working improperly.
Chapter G - TROUBLESHOOTING
CORRECTIVE ACTION
CORRECTIVE ACTION
film
CORRECTIVE ACTION
Planmeca Proline XC X-ray unit G-15