Chapter D - ADJUSTMENT & CALIBRATION
Touch the ready field to drive the X-ray unit to the exposure position. When the X-ray unit is
ready field is green.
Figure 42
Start to take a test exposure by pressing and holding down exposure button on the remote
control. Stop pressing the exposure button about halfway through the exposure cycle so that
the rotating unit stops moving.
An error code will appear on the display. Clear the error code from the display by touching the
OK field.
Manually adjust the rotating unit position so that the center line that runs through the tube
head and film cassette mechanism is parallel to the wall to which the X-ray unit is secured.
Slide the alignment pin through the hole, nearest the column, in the top of the vertical carriage
so that it goes into the positioning hole in the rotating unit.
Figure 43
NOTE
Adjust the position of the rotating unit by holding the cassette carriage. Never use the
tube head to adjust the position.
D-22 Planmeca Proline XC X-ray unit
Alignment pin
Center line parallel to wall
PATIENT POSITIONING MECHANISM
Technical Manual