Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
From the list of User settings select Behavioural preferences . The display shown below
appears. Touch the Test mode field to enable the test mode (ON). Return to the main display
by touching the OK field.
Figure 23
Press and hold down the exposure button to drive the rotating unit to a position where you
can safely view the secondary collimator when standing behind the tube head. Stop pressing
the exposure button when the rotating unit is in a convenient position.
An error code will appear on the display. Clear the error code from the display by touching the
OK field.
Exit the test mode.
Select a kilovolt and a milliampere value high enough to enable the radiation beam to be seen
in a darkened room for example 62kV and 6mA. The actual values will depend on how dark
the room is.
Touch the i-field on the Main display. The list of User preference settings appears.
Figure 24
E-12 Planmeca Proline XC X-ray unit
PANORAMIC X-RAY BEAM ADJUSTMENT