Patient Positioning Mechanism Adjustment - Planmeca Proline XC Technical Manual

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

PATIENT POSITIONING MECHANISM
4.2

Patient positioning mechanism adjustment

Adjustment in ball phantom's y-direction
Technical Manual
If the patient positioning mechanism is too near or too far from the column it will have to be
centered.
Remove the cover from the lower shelf. Replace the ball phantom tool.
Remove the sensor head from the quick connector mechanism (see instructions given in
Planmeca Proline XC panoramic X-ray unit's user's manual). Attach the sensor alignment tool
to the quick connector.
Select the Test mode as follows. Touch the i-field on the Main display. The list of User set-
tings appears.
Figure 84
From the list of User settings select Behavioural preferences . The display shown below
appears. Touch the Test mode field to enable the test mode (ON).
Figure 85
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT