PANORAMIC X-RAY BEAM ADJUSTMENT
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Make sure that the X-ray unit is in the panoramic exposure mode (text Pan on the main dis-
play). Touch the ready field to drive the X-ray unit into the exposure position. The rotating unit
will move to ready position.
Figure 21
On the exposure switch holder a ready indicator light will come on and on the main display the
Ready field will be green. The Dimaxis program will show the Waiting for Ready message on
the computer screen.
The indicator light will flash while the unit moves to the ready position. They will stop flashing
when the unit has reached the ready position for the selected panoramic exposure program.
The Dimaxis program will now show the Waiting for Exposure message on the computer
screen.
NOTE
The ready indicator light will be flashing if the PC is not in the ready state. Refer to the
Dimaxis user's manual on how to prepare the system ready for the exposure.
Select the Test mode as follows. Touch the i-field on the Main display. The list of User set-