2.5 Error history

To make the equipment failure analysis easier an error history display has been provided.
This error memory contains the last 50 errors occurred (not Er 00). The list is shown in chron-
ological order (by date/time). The oldest error messages are dropped out from the list if the
list is full and there are new errors that are stored. Browse the error list with the arrow fields.
Error code number
Time
Date
Error serial number
NOTE
The error history list can be cleared on the Factory preset window by touching the
Exposure counter button.
B-12 Planmeca Proline XC X-ray unit
The primary collimator limit sensor operation. When operated by the motor movement the
active time of the sensor is rather short and is not necessarily always visible on this indica-
tor. (Only in Planmeca Proline XC Pan/Ceph X-ray unit).
Up/down motor limit sensor
Temple support motor limit sensor
Layer adjust motor limit sensor
Rotating arm end point limit sensor
Rotating arm start point limit sensor
Cassette holder start (ready) point limit sensor
Cassette holder end (return) point limit sensor
SERVICE SETTINGS