Planmeca Proline XC Technical Manual page 232

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

Chapter H - PARTS REPLACEMENT & REPAIR
Unscrew the four Generator processor PCB attachment screws and detach the Generator
processor PCB from its position.
Figure 7
Remove the software chip from its socket with the special tool as shown on Fig. 8 below.
Figure 8
Carefully place the new software chip into the socket.
NOTE
The orientation of the software chip is critical. Never try to force the chip into the
socket.
Attach the Generator processor PCB to its position and attach cables to the PCB. Attach the
tube head inner cover and the rotating arm cover.
H-10 Planmeca Proline XC X-ray unit
REPLACING THE SOFTWARE
Technical Manual