Adjusting The Aec-Signals - Planmeca Proline XC Technical Manual

Hide thumbs Also See for Proline XC:
Table of Contents

Advertisement

AUTOMATIC EXPOSURE CONTROL (AEC) CALIBRATION

6.3 Adjusting the AEC-signals

NOTE
Technical Manual
Take five 10 seconds exposures in the service settings mode with a kilovolt value of 68
and a milliampere value of 7 before adjusting the AEC-signals.
Enter the service settings mode. From the list of service settings select Test exposure.
Figure 67
The display shown below appears. Switch ON the AEC by touching the AEC field.
Figure 68
Chapter D - ADJUSTMENT & CALIBRATION
Planmeca Proline XC X-ray unit D-39