CEPHALOSTAT ADJUSTMENT
NOTE
Calibrating the sensor head
Dimax3 cephalometric
exposure
Technical Manual
The calibration must be performed with 3x3 (enhanced resolution) and 4x4 (normal resolu-
tion) binning. With enhanced resolution select 60 kV and 11 mA, and with normal resolution
select 60 kV and 6 mA. Note that these values must be selected to produce a clear exposure.
The 2 mA value can be selected in the test exposure mode, refer to section 2.2 "Test
exposure" on page B-10.
Start the Dimax3 calibration program by double-clicking the Dimax2Tool.exe program icon
located in the Dimaxis program folder. The Dimax2Tool window appears.
Figure 150
Select Proline from the Settings / Type pulldown menu.
Figure 151
The calibration must be performed with both 3x3 (enhanced resolution) and 4x4 (normal res-
olution) binning. Select the binning from the Settings / Binning pulldown menu.
Figure 152
Click the Dimax3 cephalometric exposure button. The pmpcal2 window with text Waiting
appears.
Figure 153
Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT