Chapter D - ADJUSTMENT & CALIBRATION
Stand behind the tube head and protect yourself from radiation. Press and hold down the
exposure button. The image of the radiation beam from collimator number 6 will appear on
the alignment tool.
NOTE
The soft tissue filter will filter a narrow area of the left side of the radiation beam.
The radiation beam must appear within the borders of the rectangle marked on the alignment
tool. If it does not refer to 2.3 "Primary collimator adjustment, Planmeca Proline XC Pan/Ceph
X-ray unit" on page D-12. The adjustment is performed in the same way as when adjusting
the standard panoramic collimator.
6
Figure 92
When all the collimators have been checked, and if necessary adjusted, exit the service set-
tings mode.
D-56 Planmeca Proline XC X-ray unit
Beam
Positioning light 25mm
Positioning light 0mm
Soft tissue filter 0mm
CEPHALOSTAT ADJUSTMENT
Sixth primary collimator
Technical Manual