Chapter D - ADJUSTMENT & CALIBRATION
The radiation beam must appear within borders of the rectangle marked on the beam align-
ment tool and to the right of the line marked "Soft tissue filter 0mm".
Figure 82
If the left-hand side of the radiation beam does not reach the line marked "Soft tissue filter
0mm" or extends passed the line the soft tissue filter is in the wrong position it must be repo-
sitioned. Refer to the section 7.5 "Soft tissue filter adjustment" on page D-51.
D-50 Planmeca Proline XC X-ray unit
4
5
PLANMECA PM 2002 Proline Cephalostat 18x24
No. 652232
"Soft tissue filter 0mm" line
Area of soft tissue filter
CEPHALOSTAT ADJUSTMENT
Radiation beam
Technical Manual