Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
7.8

Second primary collimator adjustment

Second primary collimator in too low or high position
Loosen all the three screws located on the side of second primary collimator with the 4mm
allen key.
Loosen these three screws.
Figure 144
Move the second primary collimator in vertical direction until the collimator is correctly posi-
tioned. Tighten the loosened screws.
Second primary collimator not vertical
Loosen the two screws located on the side of second primary collimator.
Loosen these two screws.
Figure 145
E-78 Planmeca Proline XC X-ray unit