Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Adjust the head support angle with the two screws located on the left side of the mechanism
(use 2.5mm allen key) (see Fig. 121) so that the sensor alignment tool is perpendicular to the
cephalostat head support.
Figure 121
E-64 Planmeca Proline XC X-ray unit
CEPHALOSTAT ADJUSTMENT
90°
Spirit level
Technical Manual